Recommended safety distance
The following table lists the recommended safety distance between the device and mobile RF
communications equipment.
The recommended EM environment for the NYA Quanta Wireless Home Health Management
System is where RF interference is controlled. The user should maintain the recommended
safety distance with the devices and portable/mobile RF communications equipment listed in
the following table (corresponding to the maximum output of the RF transmitter).
Recommended Safety Distance (corresponding to RF transmitter
output)
Maximum output of
RF transmitter (Watt)
150 kHz to 80 MHz
d
=1.17
P
80 MHz to 800 MHz
d
=1.17
P
800 MHz to 2.5 GHz
d
=2.33
P
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
1 1.2 1.2 2.3
10 3.7 3.7 7.4
100 12 12 23
If the maximum output of the transmitter is not shown in the above table, use the formula to
derive the recommended safety distance
d
(Meter) with
P
(Watt) being the maximum output
of the transmitter.
Attention: For frequencies between 80 MHz and 800 MHz, use the higher frequency range.
Attention: These guidelines may not be applicable to all situations. Electromagnetic waves
can be absorbed or reflected by buildings, objects and people.