Supplier's Declaration 
The  NYA  Quanta  Home  Health  Management  System  conforms  to  the  international  EN 
60601-1  and  EN  60601-1-2  standards  for  electromagnetic  compatibility  with  medical 
electrical devices and systems. 
Emission Testing
Compliant 
Standard 
Electromagnetic Environment Guidelines 
The  recommended  electromagnetic  environment  for  this  instrument  is  shown  in  the 
table below.    The customer or user should ensure that this instrument is used in the 
described environment.   
RF Emissions 
CISPR 11 
Group 1 
Radio-frequency is only used by the internal 
functions of this instrument. The emission 
energy is also very low and does not interfere 
with nearby electronic instruments. 
RF Emissions 
CISPR 11 
Class B 
This instrument is suitable for home use 
(including residential or other mixed 
residential/public low-voltage power grids) 
Harmonic 
Emissions 
IEC 61000-3-2 
Voltage 
Fluctuations/ 
Flicker 
Emissions 
IEC 61000-3-3 
Testing 
IEC 60601-1-2 
testing 
Compliant 
Standard 
Electromagnetic 
Environment 
Guidelines 
The  recommended  electromagnetic  environment  for  this  instrument  is  shown  in  the 
table  below.    The  customer  or  user  should  ensure  that  it  is  used  in  the  described 
environment. 
Electrostatic 
Discharge 
(ESD) 
IEC 61000-4-2 
±6 kV contact 
±8 kV air 
±6 kV contact 
±8 kV air 
The floor should be 
timber, concrete or 
ceramic tile.    If the 
floor is covered in a 
synthetic material then 
it should have a 
relative humidity of at 
least 30%. 
Power Frequency 
(50/60Hz)   
Magnetic Field 
IEC 61000-4-8 
3 A/m  3 A/m 
Power frequency and 
magnetic field in 
representative 
commercial or hospital 
environments. 