User's Manual

MH8418-01 32
IEC 61000-
4-4
+ 1kV for
input/
output
lines
Not
applicable
home healthcare
environment.
Surge
IEC 61000-
4-5
+ 0.5kV,
+1kV
line(s) to
line(s)
+ 0.5kV,
+1kV,+
2kV line(s)
to earth
+ 0.5kV,
+1kV
line(s) to
line(s) Not
applicable
Mains power
quality should be
that of a typical
home healthcare
environment.
Voltage
Dips, short
interruptions
and voltage
variations
on power
supply input
lines
IEC 61000-
4-11
Voltage
dips:
0 % UT;
0,5 cycle
0 % UT; 1
cycle
70 % UT;
25/30
cycles
Voltage
interruptions:
0 % UT;
250/300
cycle
Voltage dips:
0 % UT; 0,5
cycle
0 % UT; 1
cycle
70 % UT; 25
cycles
Voltage
interruptions:
0 % UT; 250
cycle
Mains power quality
should be that of a
typical home
healthcare
environment. If the
user of the Q-temp-
w1 requires continued
operation during
power mains
interruptions, it is
recommended that
the Q-temp-w1 be
powered from an
uninterruptible power
supply or a battery.