881/882 Video Test Generator User Guide (Rev A.22) 499
2. In the ACA, from the File menu, select Connect to Analyzer. The Connect to
Analyzer dialog box appears.
3. In the Connect To box, enter the IP address of the generator you want to connect to,
and then click Connect.
4. From DDC menu, point to Port, and the select the port you want to monitor.
5. On the toolbar, click the DDC and Events buttons to enable them. Make sure all other
buttons are disabled.
6. From the File pull down menu specify New Trace.
7. Click on the Capture activation button.
8. Initiate the actions necessary to cause a hot plug event for EDID transmission to occur.
This might be unplugging the HDMI cable on the generator if the 882 is emulating a
source for testing an HDMI sink.
The ACA begins the trace and displays the results in the main window, as shown
below.