Quick Start Guide

22 882E Quick Start Guide - HDCP Compliance Testing with Encryption Status Tester (EST) - P/N 68-00210 Rev. B
10. Select Repeater3AB device by pressing the adjacent soft key to specify that the device
under test is a repeater.
The following is displayed on the generator’s LCD.
Alternatively, you can specify the device under test as a repeater for the 3A test using
the following commands:
CPTX:DUTT 3 // specifies the device under test as a repeater
for test 3AB.
You can query the current value of the parameter with:
CPTX:DUTT? // returns the device under test type.
11. (Optional) Specify the remaining parameters in the EditPCP menu using the table
above.
Alternatively, you can specify the parameters through the command line as follows:
CPTX:RPFT 0 // indicates whether DUT (repeater) supports 1.1
features such as Advanced Cipher and Enhanced
Link Verification. 1 = yes; 0 = no.
CPTX:RKSV 10 // specifies maximum number of downstream devices
that can be supported in the repeater’s KSV list
CPTX:ROOR 0 // indicates whether DUT (repeater) outputs outputs
content to the downstream repeater that does not
have any downstream device connected.
1 = yes; 0 = no.
12. Press the Options key to save the capabilities definition.
The generator LCD will display the message “Saved”.
13. Press the Tools key get back to the Reports menu.
The generator LCD will display the message “Saved”.
Repeater Max KSV Specifies the maximum number of KSVs the repeater
can read. The valid values are 2 through 127.
Repeater Out OnlyRep Indicates whether the DUT outputs content to the down-
stream repeater that has no downstream device con-
nected. The values are + for yes, the repeater will
forward encrypted video to a downstream repeater
when there are no other downstream devices; and - for
no, the repeater will not forward encrypted video to a
downstream repeater when there are no other down-
stream devices.
Parameter Explanation
Repeater3AB :DUT Type
0
1
+Source_Out_OnlyRep
:Source_Authe_Cnt
:Source_Max_KSV