HD 9021 INSTRUCTIONS MANUAL
HD 9021 1 19 2 5 4 3 21 20 22 23 6 24 7 26 9 25 8 27 11 30 10 29 13 28 12 31 14 32 15 33 16 34 17 35 18
ENGLISH HD 9021 QUANTUM - PHOTO - RADIOMETER – 25 –
ENGLISH 1 Input A socket, DIN 8-pole connector 2 Display 3 REL: the symbol indicates that the instrument is storing the Maximum, Minimum and Q energy values 4 MAX: the symbol indicates the maximum value of the chosen unit 5 MIN: the symbol indicates the minimum value of the chosen unit 6 A, B, A-B: the symbol indicates the relative value at the input A, B, or the difference between the two inputs A-B 7 AUTO: the symbol indicates that the automatic change of scale device is enabled 8 °C/°F: key for selectin
ENGLISH Key Symbols lit besides the numbers Description All the symbols are lit for a few seconds after switching on. Key for switching the instrument on and off. The instrument switches off automatically about 8 minutes after this key has been pressed, as it has a built-in switch-off device. If any key other than the ON/OFF key is pressed, the instrument switches off 8 minutes after the last key was pressed.
ENGLISH Key Symbols lit besides the numbers Description shes. During the RCD phase the AUTOPOWER-OFF function is disabled, that is the instrument does not switch off automatically. To switch off, press the ON/OFF key. If the stored data are not reset they remain in the memory until REC is pressed again.
ENGLISH Symbols lit besides the numbers Key Description When the MAN key is pressed the user chooses what scale to operate on. There are 4 scales available, except for temperature measurement where there are 2. s m h When this key is pressed the Q Energy integration time that has elapsed since the REC key was pressed appears on the display in seconds, minutes or hours. To quit the routine, press the key for the channel in which the reading is being taken.
ENGLISH Symbols lit besides the numbers Key Description if a high value is to be set. This key has two functions; in the first, when Q is set, it increases the value; the second function enables the serial output. When the key is pressed only once, the figure shown on the display is sent to the serial output. If the key is pressed for more than three seconds the continuous serial output is enabled; this means that what is shown on the display is sent continuously to the serial output every 10 seconds.
ENGLISH Symbols lit besides the numbers Key Description Err If the probe is not connected to the instrument when it is switched on, or if there is a break in the probe, the "Err" message appears. E1 This message appears on the display when the user tries to obtain the difference between inputs A and B when two probes of different types are connected. OFL This symbol appears on the display when the reading is out of scale.
ENGLISH The light emitted by a lamp varies with the square of the current, so there may very well be variations in values during the test which the human eye does not detect. When measuring temperature choose the probe most suited for the case: immersion, penetration or surface probes.
ENGLISH LOW BATTERY WARNING If the battery voltage falls below acceptable levels, a beep sounds every 30 seconds and the H symbol appears. From that moment there remains about 1 hour autonomous operation, however the battery should be replaced as soon as possible. To change the battery turn the screw on the door of the battery compartment in an anti-clockwise direction.
ENGLISH MAINTENANCE Storage conditions. -20 to +60°C * Temperature: less than 85% relative humidity. * Humidity: store the instrument in places where: * Do not 1 - There is a high degree of humidity. 2 - The instrument is exposed to direct sunlight. 3 - The instrument is exposed to a source of high temperature. 4 - There are strong vibrations. 5 - There is steam, salt and/or corrosive gas. The instrument body is made of plastic so it must not be cleaned with solvents which can spoil plastic.
ENGLISH DTE DCE AD RS 232C 13 12 11 10 9 8 7 6 5 3 2 NC NC 10 11 23 12 1 5 4 25 24 23 22 21 20 19 18 17 16 15 14 13 24 25 NC 22 TEST NC DTR 9 21 NC 8 20 NC NC NC 7 19 NC 6 18 4 DCD DSR 5 17 NC 4 16 GND CTS RTS TC NC 3 15 NC 2 14 NC 1 RC TXD RXD SCR DTE - Data terminal equipment COMPUTER DCE - Data communication equipment MODEM 9 3 8 2 7 1 6 ANALOG OUTPUT It is possible to link up with the SUB D male 9-pole connector of the instrument to obtain an a
– 36 – INPUT B CONNECTOR (Range) Signal Probe Type (Range) Probe Type Signal INPUT A CONNECTOR POWER SUPPLY 4V TR1 500 mV TR2 A/D CONVERTER KEYBOARD RS232 MICROPROCESSOR LOW VOLTAGE DETECTOR LCD ENGLISH BLOCK DIAGRAM
C2 U1 C3 DZ1 – 37 – R35 R36 Q1 C1 R3 R5 R4 R25 TR1 C7 R29 G2 R11 C14 R10 Q4 C16 R4C6 R24 D1 R8 R39 R6 C4 R6 U3 TR2 R23 R33 R34 OUTA B2 R23 R33 R34 R9 OUTB G3 C5 U4 ICON1 J5 U2 C9 C10 C11 R45-51 C8 R30 R28 R32 C15 R26 R2 R1 C13 C12 R27 X1 R41 U5 R38 R18 R17 R37 R15 R16 TR1 R14 R12 TR2 R20 R19 R40 J4 J3 ENGLISH LAYOUT OF COMPONENTS
ENGLISH CALIBRATING THE INSTRUMENT HD 9021 1) Open the instrument. 2) Connect the prods of a precision meter to the following pins on the input A (or B) connector: the positive to pin 7 (V+) and the negative to pin 6 (GND). 3) With the instrument and the meter switched on, turn the trimmer TR1 until the display on the meter indicates 4Vdc. 4) Insert the photometric simulator in the connector A (or B). 5) Turn the trimmer TR2 until the display on the instrument shows the same value as the simulator.
ENGLISH CHARACTERISTICS - Microprocessor-controlled multifunction quantum-photo-radiometric indicator with LCD indication for measuring sources of light and temperature. Measures ILLUMINANCE (lux, foot-candle), IRRADIANCE (W/m2) and RADIANCE (cd/m2). - 2 inputs for probes and one for RS232C (with an optional adapter AD RS232C). - Probe connection: 2 DIN 45236 8-pole sockets. - Different silicon sensors for various types of measurement.
ENGLISH - Display: LCD 12 mm, function and photo-radiometric symbols. - Functions: Automatic and manual change of scale, integration time up to 1999 hours, auto power off. Storage and updating of the Maximum, Minimum and Q (Energy) values. Optoinsulated serial output RS232C with adapter AD RS232C. Analog output 1.9 mV/°C, 2.5 mV/lux, 250 mV/W. - Power supply: 9V battery IEC 6LF22. - Battery life: 150 hours with alkaline battery.
ENGLISH ORDER CODE HD 9021 K : The kit is composed of the HD 9021 instrument, diplomatic carrying case, without probe. HD 9021 : Photometric probe for measuring light, illuminance, photopic PHOT/C correction filter complying with CIE, diffuser for cosine correction. Range 0÷200,000 lux. HD 9021 : Radiometric probe for measuring light energy, irradiance, RAD/C radiometric filter, diffuser for cosine correction. Range 400÷900 nm.
ENGLISH PROBES The primary purpose of the probes is to detect and measure light in its various aspects; for this reason there are different probes available to suit the different requirements. Probes are required for measuring illuminance (lux, foot-candle), irradiance (W/m2) and luminance (cd/m2). Each sensor has its own work band completed by filters or diffusors which correct or limit its range. The size of the silicon sensor used in the various probes is 7.34 mm2.
ENGLISH HD 9021 RAD/C Radiometric probe for measuring the IRRADIANCE of light sources, irradiance of the sun, etc. Radiometric filter and diffuser for angle correction according to the cosine law. Spectral measurement range 400÷900 nm, range from 1 microwatt/cm2 to 200 milliwatt/cm2. The probe is used in greenhouses for measuring the power of artificial light, shade, the performance of solar panels, solar heating plants, solar filters, etc. The typical response curve of the probe is as follows: 1.00 0.
ENGLISH The typical response curve of the probe is as follows: 1.00 0.89 0.78 0.67 0.56 0.44 0.33 0.22 0.11 900.0 850.0 800.0 750.0 700.0 650.0 600.0 550.0 500.5 450.0 400.0 0.00 The study and field of application of ultraviolet rays has seen considerable growth in the last decades. The ultraviolet field is conventionally subdivided into the following three "bands": - the band between the wavelengths from 400 to 315 nm.
ENGLISH UVC rays are useful thanks to their bactericide and ozonizing capacity, so they are used in the field of sterilization; another field of application that is being developed is in the polymerization of adhesives or resins (photoresist); as a drawback they cause more serious cases of erythema and conjunctivitis than UVB rays.
ENGLISH HD 9021 UVB Radiometric probe for measuring radiation power, IRRADIANCE, in the UVB wave length 280÷315 nm, peak at 312 nm. Quartz diffuser for cosine correction. Measuring range from 10 nanowatt/cm2 to 200 milliwatt/cm2. The typical response curve of the probe is as follows: 1.00 0.90 0.80 0.70 0.60 0.50 0.40 0.30 0.20 0.10 400.0 380.0 360.0 340.0 320.0 300.0 280.0 260.0 240.0 220.0 200.0 0.
HD 9021/Cd Probe for measuring LUMINANCE, measuring range from 0 to 1999x103 candles/cm2. Measuring angle 6° with opening surface 25 mm. CIE filter for correction according to the human eye. 6° - The CANDLE (cd) is the light intensity in a specific direction of a source that emits monochrome radiation at a frequency of 540x1012 Hz and whose energetic intensity in that direction is 1/683 W/sr.
ENGLISH CE CONFORMITY Safety Electrostatic discharge Electric fast transients Voltage variations Electromagnetic interference sucseptibility Electromagnetic interference emission EN61000-4-2, EN61010-1 level 3 EN61000-4-2 level 3 EN61000-4-4 level 3 EN61000-4-11 IEC1000-4-3 EN55020 class B GUARANTEE CONDITIONS All our appliances have been subjected to strict tests and are guaranteed for 24 months from date of purchase.