User's Manual

SAMPLE LOADING AND TESTING / REPEATING TESTING BY “DATA REDUCTION ERROR”
105
N
ERR
O
R”
MosaiQ 125 - User Manual - November 2020 - Reference n°: MMSURM.0010.02
8.13 REPEATING TESTING BY “DATA REDUCTION ERROR”
When a result is marked as “Data Reduction Error, the test has been failed and no results can be
generated.
INFO
On each microarray there are several control probes showing that
the assay is reacting as expected. All control probes must have an
appriopriate reaction for the test to be interpreted by the MosaiQ
Software. The failure of one or more of these control probes during
the image analysis will provide a “Data Reduction Error” message
instead, meaning that no result can be generated.
To check the details of the test impacted
-
1
Tap on the Results icon in the navigation bar.
-
2
Tap on the Review sub menu.
-
3
Tap on the desired test result to show the details of the results.
3
1
2
(review screen)
(test results report in collapsed view)
-
4
Tap on the test result message or on Review button to decide if you want to repeat the test.
4
(test results review)