User's Manual

163
MosaiQ 125 - User Manual - November 2020 - Reference n°: MMSURM.0010.02
15. INSTALLATION AND REMOVAL OF THE MOSAIQ 125
15.1 MOSAIQ125 INSTALLATION
CAUTION
The installation of the MosaiQ 125 and its connection to
external devices (e.g. printer) must be executed by Quotient
Suisse SA authorized service personnel only.
Improper installation can cause damage or malfunctions.
After the installation is completed, the user receives the qualification documentation (IQ,
OQ) attesting the installation of the MosaiQ125 has been performed properly.
INFO
Installation Qualification (IQ) verifies that the MosaiQ 125 has
been installed and configured according to the manufacturer’s
specifications or installation checklist.
Operational qualification (OQ) is performed after meeting each
protocol of IQ. OQs purpose is to determine that the MosaiQ 125
performance is consistent with the user requirement specification
within the manufacturer specified operating ranges. In action, this
means identifying and inspecting equipment features that can
impact final product quality.
Please note, the enclosed Microsoft software license terms for the Microsoft Windows
embedded operating system.
INFO
Microsoft software license terms EULA (End-User License
Agreement). The EULA specifies in detail the rights and restrictions
which apply to the use of the software.
15.2 REMOVAL OF THE MOSAIQ125
CAUTION
Removal must be executed by Quotient Suisse SA
authorized service personnel only.
Improper removal can cause damage or malfunctions.
After a displacement (within the laboraty) of the MosaiQ 125, a
re-installation is required. If omitted, this can cause damage or
irregular pipetting performance of the MosaiQ 125.