User's Manual

SAFETY INSTRUCTIONS / ELECTRICAL SAFETY
24
MosaiQ 125 - User Manual - November 2020 - Reference n°: MMSURM.0010.02
Sharp Edges
WARNING
Sheet metal parts and circuit boards located behind protective
covers may have sharp edges. Contact may lead to injuries.
Use caution at corners and edges.
Unauthorized Changes to the MosaiQ125
WARNING
Any changes to the MosaiQ 125 that are not authorized by the
manufacturer will lead to non-conformity with respect to the
applicable regulations the manufacturer has declared. In this
case, the customer is responsible for fulfilling the applicable
regulations.
Do not make unauthorized changes to the MosaiQ125.
4.2 ELECTRICAL SAFETY
For the overall safety of laboratory personnel and equipment, please read the following
information regarding electrical safety precautions before operating the MosaiQ125.
MosaiQ 125 complies with the following electrical safety standards:
IEC 61010-1
IEC 61010-2-010
IEC 61010-2-101
Not Following Rules and Regulations
DANGER
Not following rules and regulations may cause serious personal
injury with deadly consequences and material damage.
National rules and legal regulations for the safe electrical operation of the
MosaiQ125 must be observed.
Improper Connection of Mains Power Supply
DANGER
Improper connection of the MosaiQ 125, and peripheral devices
to the mains power supply can cause serious personal injury with
potential deadly consequences and material damage (e.g. fire).
Only use grounded connectors and power cords with sufficient capacity (voltage
and current) to connect the MosaiQ125 and PC to the mains power supply.
Never remove ground connections.
The use of a multi-outlet power strip is prohibited.
Only use power cords that are provided with the MosaiQ125.
Damaged Power cords
DANGER
Damaged power cords may cause serious personal injury with
potentially deadly consequences and material damage (e.g. fire).
Damaged power cords must be replaced immediately,
No objects may be placed on the power cords,
Power cords must be installed so that they cannot be squeezed or damaged,
Power cords must be laid so that they do not lay in accessible areas.
Electric Shock by Electrical Devices on Wet Surfaces
DANGER
Working with electrical devices on wet surfaces (floors, worktable)
may cause serious injuries with deadly consequences and material
damage due to electric shock.
Only work on dry surfaces (floors, worktable).
Emergency Shutdown in Case of Malfunction
DANGER
Malfunction of the MosaiQ125 may cause electrical shock, burns,
cuts or contusions.
Use the main switch to switch off the MosaiQ125 or the mains plug to disconnect
the MosaiQ125 from the mains power supply.