User's Manual

MOSAIQ125 PARTS DESCRIPTION AND LOCATION / WATER DRAWER
38
MosaiQ 125 - User Manual - November 2020 - Reference n°: MMSURM.0010.02
5.9 WATER DRAWER
The water drawer is located on the bottom right side of the MosaiQ125 and houses a
maximum of two canisters containing the system liquid required to operate the MosaiQ125.
Users can open and close the water drawer manually. The water drawer can be opened
and closed while the MosaiQ125 is in operation.
(water drawer localization)
Functions:
To store water canisters (blue caps).
To access the computer workstation.
NOTICE
Drawers
The pulled-out drawer can support the weight of system liquid
filled canisters and no additional weight
Do not lean on the drawers,
Place a maximum of two filled canisters in a drawer.
5.9.1 WATER CANISTERS
There are two canisters with a capacity of 10 L each (provided with your MosaiQ 125). They
shall be filled with purified water (also called system liquid in this manual). Both canisters
are connected to an intermediate container with a 5 L capacity.
NOTICE
Purified water is used in general laboratory applications such
as buffers, pH solutions and microbiological culture media
preparation.
(opened water drawer)
To remove and load the water canisters, please refer to chapter 7.3.11 of this manual.
To clean the water canisters (monthly task), please refer to chapter 12.7.5 and
chapter 12.7.6 of this manual.
Functions:
The water canisters store the system liquid used to prepare the reconstituted wash
buffers.
NOTICE
MosaiQ reconstituted wash buffers.
The reconstituted wash buffers are composed as follow:
10% MosaiQ Wash Buffer 1 concentration in system liquid.
10% MosaiQ Wash Buffer 2 concentration in system liquid.
Each MosaiQ wash buffer is diluted with system liquid in a dilution
chamber as described in
chapter 5.8.
The level of concentration is monitored by a conductivity sensor.
The conductivity values are displayed each time the reagent
drawer is closed.