OPERATING INSTRUCTIONS / HOW TO REMOVE AND LOAD EACH RESOURCE
71
E
MosaiQ 125 - User Manual - November 2020 - Reference n°: MMSURM.0010.02
7.3.4 LOADING DETECTION AND ENHANCEMENT REAGENT
BOTTLES
This section provides step-by-step instructions on how to load reagent bottles (MosaiQ
detection reagents and MosaiQ enhancement reagents) into the MosaiQ125. The system
software controls reagent door locking and unlocking operations.
-
1
Tap on the Resources icon in the navigation bar.
-
2
Tap on the Reagents area to access the Reagents screen.
2
1
(reagents resources area)
-
3
Wait until the MosaiQ125 status is Idle.
3
(reagents resources detais screen)
NOTICE
To know the remaining time before MosaiQ125 goes to Idle, go to
the Results menu and tap on the Processing button.