OPERATING INSTRUCTIONS / HOW TO REMOVE AND LOAD EACH RESOURCE
76
E
MosaiQ 125 - User Manual - November 2020 - Reference n°: MMSURM.0010.02
7.3.6 LOADING THE MOSAIQ WASH BUFFER AND MOSAIQ
SAMPLE DILUENT BOTTLES
This section explains how to load the bottles into the MosaiQ125.
The bottles are located in the reagent drawer.
The reagent drawer can be opened or closed when the MosaiQ125 is in operation.
1 - Open the drawer.
(opening the reagent drawer)
2 - Insert the MosaiQ wash buffer and MosaiQ sample diluent bottles upside down in the
appropriate location.
3 - Press the bottle down slightly.
(inserting the buffer and diluent bottles)
4 - Bottles can be loaded in duplicate for higher throughput if required.
1
2
3
3
4
(2 slots for each MosaiQ wash buffer and MosaiQ sample diluent bottles)