7.3.9 REMOVING LIQUID WASTE CANISTER
This section explains how to remove the liquid waste canister from the MosaiQ125.
The liquid waste canister is located in the waste drawer. Even when the MosaiQ125 is in
operation, it is possible to remove one of the cannisters even if its not completely full.
The liquid waste drawer can be opened and closed manually.
INFO
Re-usable canister
The liquid waste canister is re-usable and can hold up to 10 L.
The liquid waste canister must be cleaned monthly, please see:
chapter 12.7.1 of this manual.
CAUTION
The liquid waste canister should be filled in accordance with the
laboratory local manual handling guidelines for maximum weight
limit at work. It is recommended to make a visual mark on the
canister to visualize the filling limit complying with the maximum
weight as per local regulation.
When filled with 10 L of liquid waste, a canister can weight up to
11,2 kg.
-
1
Tap on the Resources icon in the navigation bar.
-
2
Tap on the Liquid Waste / Solid Waste area to check the level.
1
2
Shows the overall
fill level of the liquid
waste containers as
a percentage.
Shows the quantity of the
solid waste (microarrays)
(waste capacity area)