User's Manual

Table Of Contents
Series 6.0 Software (0510-1065-A) - User Guide 81
Event Types
Link Alarm
This alarm occurs when there is a problem linking the Infant transmitter to a Mother transmitter. For example,
the Infant transmitter may already be linked to another Mother transmitter.
To respond to a Link Alarm
1. Always follow your facilitys policies and procedures to ensure patient safety and secure the area.
2. Click anywhere in the Link Alarm Message Box to access the Event Information window.
3. Click Clear.
4. If necessary, verify the identity of the mother and infant then relink the Mother and Infant transmitters.
Check Alarm (not “Check Transmitter Alarm”)
No message has been received from the patients transmitter or the device in the patient’s room during the
Inactivity Check-In period. The Check message will remain until the patient’s transmitter checks in, an Alarm
message is received, or the event is manually cleared at the Client computer(s).
To respond to a Check Alarm
1. Always follow your facilitys policies and procedures to ensure patient safety and secure the area.
2. Locate the device.
3. Verify that no patient emergency exists.
4. From the Client computer, click anywhere in the Inactivity Check-In Alarm Message Box to access the
Event Information window.
5. Select Clear to clear the Inactivity Check-In alarm.
NOTE: You can click Silence to stop the alarm sound at the Central Server or Client
computer. The alarm is silenced at that computer only, for the configured length of time;
however, the event still appears in the Event List. The next event automatically restarts the
alarm sound.