Series 6.0 Software (0510-1065-A) - User Guide 85
Event Types
Wet Alarm
This alarm occurs when fluids are detected on an incontinence pad.
To respond to an Wet Alarm
1. Always follow your facilitys policies and procedures to ensure patient safety and secure the area.
2. Proceed to the patient and verify incontinence occurrence. Detach the incontinence pad; the alarm
clears when the incontinence pad is detached.
3. If the Enforce JCAHO feature is activated, you must select an Event Cause once the alarming device
has been reset. When you reset the alarming device, the Red Alarm changes to a White Alarm in the
Alarm Message Box and Event List. If JCAHO is not activated, the Alarm Message Box clears once the
alarm is cleared at the device.
4. From the Client computer, click anywhere in the Wet Alarm Message Box to access the Event
Information window.
5. Select one of the following event causes:
Urinethe patient has urined on him/herself.
Sweat—the patient has sweated causing the bed to be wet.
Stool—the patient has defecated on him/herself.
Spill—the patient has spilled liquid on the bed.
Other—opens a Clear Reason window, user must enter the cause in this window.
You cannot clear an event until you select the appropriate cause for the event. The event is cleared from the Event
List of every Client computer configured to monitor the unit.
NOTE: You can click Silence to stop the alarm sound at the Central Server or Client
computer. The alarm is silenced at that computer only, for the configured length of time;
however, the event still appears in the Event List. The next event automatically restarts the
alarm sound.