Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems ArbConnection 4-27
output.
To access the required parameter, click on the button below
parameters sub-
group until the LED next to the required parameter
turns on. The value that is associated
with the lit LED is displayed
on the digital display. You can use the dial, keyboard, or the [] [}
keys to adjust the readout to the required setting. After you modify
the reading, click on the Modify/Execute knob to update the 3152B
with the new reading.
ASK
The ASK group contains parameters that control the amplitude shift
keying function. These are: control data, non-
modulated and shifted
phases, and baud and marker position.
Control Data
The Control Data button in the ASK group provides
access to
the data string that controls the sequence of base
amplitude and shifted amplitude. It contains the list of0” and “1”
values that the output will repeatedly follow for its amplitude shift
keying sequence advance as programmed.
“0/1” Amplitude In AS
K, the carrier waveform (CW) has two
amplitudes: an initial amplitude level which is set by the “0”
Amplitude parameter and shifted amplitude which is set by the “1”
Amplitude. The control data table has a list of “0” and “1” values that
flag when the amplitude shifts from base to shifted amplitudes.
Baud The baud parameter sets the rate at which
the generator
steps through the sequence of the ASK Control Data bits.
Marker Index
The marker setting programs a step (index) in the
control data string to ou
tput a pulse at the SYNC output connector.
The SYNC State button must be turned on to generate the ASK
marker output.
Ampl/Freq Hop
The Ampl/Freq Hop panel (Figure 4-18), contains parameters for
controlling the hop modulation function. To turn one of the
functions on and off, click on the appropriate button in the State
group. The output has two hop options: Fixed and Variable. In
Fixed mode, the output steps through the pre-assigned hop values
at a constant rate, as programmed using the dwell time parameter.
In Variable mode, the output dwells on each step for a period of
time that is programmed in the Dwell Time field in the hop data
table that is programmed when using the Variable Hold option.
The various groups in this panel are described below.
State
The State buttons enable or disable the hop functions.
General
The General group contains parameters that are common to
all of
the hop functions. These are CW frequency and baseline.
CW Frequency The CW Frequency is the frequency of the pre-