User manual

Table Of Contents
3152B User Manual Publication No. 980935 Rev. M
5-4 Programming Reference Astronics Test Systems
decimal points, and scientific notation. Special values for numeric
parameters like MINimum and MAXimum are also accepted.
Engineering unit suffi
x
es with numeric parameters (e.g., MHz or kHz)
can also be sent. If only specific numeric
values are accepted, the
function generator will ignore values, which are not allowed and will
generate an error message. The following command is an example of
a command that uses a numeric parameter:
VOLT:AMPL <amplitude>
Discrete Parameters
Discrete parameters are used to program settings that have a limited
number of values (i.e., FIXed, USER and SEQuence). They have
short and long form command keywords. Upper and lowercase letters
can be mixed. Query responses always return the short form in all
uppercase letters. The following command uses discrete parameters:
SOUR:FUNC:MODE {FIXed | USER | SEQuence}
Boolean Parameters
Boolean parameters represent a single binary condition that is either
true or false. The generator accepts "OFF" or "0" for a false condition.
The generator accepts "ON" or "1" for a true condition. The instrument
always returns "0" or "1" when a boolean setting is queried. The
following command uses a boolean parameter:
OUTP:FILT { OFF | ON }
The same command can
also be written as follows:
OUTP:FILT {0 | 1 }
Arbitrary Block
Parameters
Arbitrary block parameters are used for loading waveforms into the
generator's memory. Depending on which option is installed, the
3152B
can accept binary blocks up to 1M bytes. T
he following
command uses an arbitrary block parameter that is loaded as binary
data:
TRAC:DATA#564000<binary_block>
Binary Block
Parameters
Binary block parameters are used for loading segment and sequence
tables into the generator's memory. Information
on the binary block
parameters is given later in this manual.
SCPI Syntax and
Styles
Where possible the syntax and styles used in this section follow those
defined by the SCPI consortium. The commands on the following
pages are broken into three columns; the Keyword, the Parameter
Form, Default and HS command equivalent.
The Keyword column provides the name of the command. The actual
command consists of one or more keywords since SCPI commands
are based on a hierarchical structure, also known as the tree system.
Square brackets ( [ ] ) are used to enclose a
keyword
that is optional
when programming the command; that is, the 3152B will process the