3152B User Manual Publication No. 980935 Rev. M
5-36 Programming Reference Astronics Test Systems
the offset + amplitude does not exceed 16.
<MINimum>
Discrete
Set
s the amplitude to the lowest possible level (10mV).
MAXimum>
Discrete
Sets
the amplitude to the highest possible level (16V).
Response
The
3152B returns the current amplitude value. The returned value will be in scientific notation, e.g., 100 mV
would be 100e-3.
VOLTage:OFFSet <offs>(?)
HS Syntax: 0501A<float>(?)
Description
Programs
the amplitude offset of the output waveform. The offset is calibrated when the source impedance is
50
.
Parameters
Name
Range
Type
Default
Description
<offs>
-
7.992 to
7.992
Numeric
0
Sets
the offset of the output waveform in units of Volts
.
Offset and amplitude settings are independent providing
that the offset + amplitude do not exceed the specified
window.
Response
The
3152B returns the current offset value. The returned value will be in scientific notation, e.g., 100 mV
would be 100e-3.
PHASe:OFFSet <phase_offs>(?)
HS Syntax: 0027A<float>(?)
Description
This command will affect a slave instrument only when
it is synchronized to another module in the chassis. It
programs the start phase offset
with respect to an adjacent master module. Phase offset resolution when
using this command is 1 point.
Parameters
Name
Range
Type
Default
Description
<phase_offs>
0 to 1e6
-1
Numeric
(Integer
only)
0
Sets
the phase
offset in reference to a master
instrument. Slave instruments trail the master
instrument edge. The range is extended 4
Meg-
1 when
the 4 Meg memory option is installed
. 1 Meg
is
standard.
Response
The 3152B returns the current phase offset value.
FUNCtion:MODE {FIXed|USER|SEQuence|SWEep|MODulation|
COUNter
|DPULse|HALFcycle}(?)
HS Syntax:
0610A<0|1|2|3|4|5|6|7>(?)
Description
This command selects the operating mode of the 3152B.
Parameters