User manual

Table Of Contents
Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems Programming Reference 5-51
be loaded into the instrument's memory. Correct memory
management is required for best utilization of the arbitrary memory.
An explanation of how to manage the arbitrary waveform memory is
given in the following paragraphs.
Arbitrary memory Management
The arbitrary memory in comprised of a finite length of words. The
maximum size arbitrary waveform that can be loaded into memory
depends on the option that is installed in your instrument. The various
options are listed in Chapter 1 of this manual. If you purchased the
3152B
with in its basic configuration then you have 1 Meg words
for
the
loading of arbitrary waveforms.
Waveforms are cre
ated using small sections of
arbitrary memory. The
memory can be partitioned into smaller segments (up to 16k) and
different waveforms can be loaded into each segment, each having a
unique length. Minimum segment size is 16 points. Information on
how to partition the memory, define segment length
,
and download
waveform data to the 3152B is given in the following paragraphs.
Table 5-6, Arbitrary Waveform Command Summary
Keyword
Parameter Form
Default
3152A
HS
:FORMat
:WAVE
NORMal | USER
Error, not supported
0101A
:RESolution
16BIT | 12BIT
16BIT
0102A
:INSTrument
LEGacy | MODern
MOD
0189A
:BORDer
NORMal | SWAPped
NORM
0100A
:TRACe
[:DATA]
<data_array>
10001#
:DEFine
<1 to 10k>,<16 to 1(2/4)e6> (<segment_#>,<size>)
10102
:DELete
[:NAME]
1 to 10k
09201
:ALL
10210
:SELect
1 to 10k
1
1030A
:SEGMent
[:DATA]
<data_array>
10501#
FORMat:WAVE {NORMal|USER}(?)
HS Syntax:
0101A<0|1>?
Response
This 3152A command is not supported by the 3152B. The 3152B returns an error.