Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems Performance Checks
6-35
Test Results Pass Fail
PSK
Equipment: Oscilloscope
Preparation:
1. Configure the oscilloscope as follows:
Time Base: 0.2 ms
Amplitude: 1 V/div.
2. Connect the 3152B output to the oscilloscope input, channel 1
3. Connect the 3152B SYNC to the oscilloscope input, channel 2
4. Configure model 3152B controls as follows:
Reset
Waveform: Modulated
Modulation: PSK
Carrier Freq: 10 kHz
Shift Phase: 180°
Baud Rate: 10 kHz
Sync: On
Output: On
5. Using ArbConnection, prepare and download 10-step PSK
list with alternating “0” and “1”
Test Procedure:
1. Verify PSK operation on the oscilloscope as follows:
Waveform: Sine wave
Period: 0.1 ms
Phase: Every 0.1 ms change 180 degrees
Test Results Pass Fail
A
SK
Equipment: Oscilloscope
Preparation:
1. Configure the oscilloscope as follows:
Time Base: 0.2 ms
Amplitude: 1 V/div.
2. Connect the 3152B output to the oscilloscope input, channel 1
3. Connect the 3152B SYNC to the oscilloscope input, channel 2
4. Configure model 3152B controls as follows:
Reset
Waveform: Modulated
Modulation: ASK
Carrier Freq: 10 kHz
Base Amplitude: 4 V
Shift Amplitude: 2 V
Baud Rate: 10 kHz
Sync: On