Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems Adjustments and Firmware Update 7-9
Item in
Cal Panel
DMM
Function
Low Limit
(Volts)
Target
(Volts)
High Limit
(Volts)
27
VAC
-2.010
-2.00000
-1.990
Change DMM function to AC Volts
28
VAC
7.00107
7.07107
7.14107
29
VAC
3.49553
3.53553
3.57553
30
VAC
0.70111
0.70711
0.71311
31
VAC
0.34955
0.35355
0.35755
32
VAC
0.06971
0.07071
0.07171
33
VAC
0.03436
0.03536
0.03636
34
VAC
0.70111
0.70711
0.71311
35
VAC
0.34955
0.35355
0.35755
36
VAC
0.06971
0.07071
0.07171
37
VAC
0.03436
0.03536
0.03636
38
VAC
2.10132
2.12132
2.14132
39
VAC
0.34955
0.35355
0.35755
40
VAC
0.06971
0.07071
0.07171
41
VAC
0.03436
0.03536
0.03636
42
VAC
7.00107
7.07107
7.14107
43
VAC
3.49553
3.53553
3.57553
44
VAC
0.70111
0.70711
0.71311
45
VAC
0.34955
0.35355
0.35755
46
VAC
0.06971
0.07071
0.07171
47
VAC
0.03136
0.03536
0.03936
48
VAC
0.70111
0.70711
0.71311
49
VAC
0.34955
0.35355
0.35755
50
VAC
0.06971
0.07071
0.07171
51
VAC
0.03436
0.03536
0.03636
52
VAC
2.10132
2.12132
2.14132
53
VAC
0.34955
0.35355
0.35755
54
VAC
0.06971
0.07071
0.07171
55
VAC
0.03436
0.03536
0.03636
Note
After making calibration adjustments, it is essential
that you lock the calibration factors as described
below. This stores the calibration factors in flash
memory
so that the instrument will retain its
calibration when power is turned off.