User manual

Table Of Contents
3152B User Manual Publication No. 980935 Rev. M
A-10
3152B Specifications Astronics Test Systems
Counter/Timer Characteristics
Operation The 3152B has a special mode where the
instrument type is transformed to operate as a
counter/timer. When this mode is selected, the
operation of the arbitrary waveform and its outputs
are disabled
Measurement Functions Frequency, Period, Period Averaged, Pulse Width
and Totalize
Frequency, Period Averaged
Frequency Range 20 Hz to 100 MHz (typically >120 MHz)
Period Averaged Range 10 ns to 50 ms
Resolution 7 digits in one second of gate time, reduced
proportionally with lower gate times
Period, Pulse Width
Range 500 ns to 50 ms
Resolution 100 ns
Totalize
Frequency Range 20 Hz to 100 MHz
Accumulation Range 10
12
-1
General
Input Front panel TRIG IN, BNC connector
Trigger Level Range ±10 V
Sensitivity 500 mVp-p
Damage Level ±12 V
Minimum Pulse Width ≥10 ns
Slope Positive/Negative transitions, selectable
Gate Time 100 µs to 1 s
Reading Modes
Repetitive Continuous measurements are executed when
signal is present at the input
Hold Single measurement is executed upon command
Gated Active in Gated Totalize mode only
Time Base
Type TCXO
Temperature Stability 1 ppm, 0°C - 40°C
Long Term Stability 1 ppm, 1 year
Backplane Multiple Instrument Synchronization
Description Multiple instruments can be daisy-chained and
synchronized to provide multi-channel systems.
Phase (leading edge) offset between master and
slave units is programmable.
Initial Skew <± 75 ns + 4 sample clock cycle
Synchronized Waveforms Standard, Arbitrary and Sequenced using the
automatic sequence advance mode only
Synchronized Run Modes Continuous, Triggered, Gated and Counted Burst