Publication No. 980935 Rev. M 3152B User Manual
Astronics Test Systems Operation 3-3
be double the programmed value, just as it was when running the
software with the 3152A, but the accuracy will be improved.
Input/Output
Protection
The 3152B provides protection for its internal circuitry at the
input
and output connectors. Appendix A specifies
the level of protection
for each input or output connector.
Power On/Reset
Defaults
At power-up or as a result of a software reset, the
3152B changes
all settings to their default values. Chapter 4 lists all settings and
their default values, as well as their maximum and minimum
allowable values.
It is good practice to reset the instrument between sequence events
to make sure that parameters that were programmed for previous
tests will not accidently interfere with future tests.
Use the following
common command to reset the instrument to its default state:
*rst
Turning the
Output
On
For safety, the “OUTPUT” connector
of the 3152B defaults to the
“Off” state when the unit is first powered up or receives a
software
reset.
To avoid the production of random signals at the output
connector, keep the output in the “Off state
while setting up the
instrument.
The output signal is connected to the OUTPUT
connector through
a mechanical relay
. When writing software to control the 3152B,
provide a 2 ms delay to make sure that the signal output
is stable
before you proceed with the next step of the
test. Use the following
commands to turn the output on and off.
outp 1 Turns the output on
outp 0 Turns the output off
The 3152B defaults to a 1 MHz, 5 Vp-
p sine wave when you apply
power or reset the unit. If you turn the output
on before changing
any settings, the output defaults to a sine wave signal.
Turning the SYNC
Output On
For safety, the SYNC OUT”
connector of the 3152B defaults to the
“Off” state when the unit is first powered up or receives a software
reset.
To avoid the production of random signals at the output
connector, keep the SYNC output in the “Off”
state while setting up
the instrument.
The SYNC output signal is hard-wired to the output connector and,