User guide
Quantity One User Guide
6-10
The first time you use the GS-710, you must select some calibration settings to 
ensure that your calibration is accurate.
To set the automatic calibration settings, click on the More Options button in 
the GS-710 acquisition window. This will open the Densitometer Options 
dialog box.
Fig. 6-9. Densitometer Options dialog box.
6.5.a Editing the Step Tablet
Before you scan for the first time, you must enter the step tablet values for 
your transmissive calibration strip into the Step Tablet Values form.
Note: In reflective scanning, the scanner is calibrated against the reflective step 
tablet. However, you do not need to enter the values for this step tablet; 
you can use the default values in the Step Tablet Values form.
Attached to the outside of your GS-710, you will find a copy of the 
manufacturer’s printout of the diffuse density values for each segment of 
your transmissive step tablet, as well as a serial number that identifies the 










