User guide
GS-710 Imaging Densitometer
6-13
Diffuse density values are converted to specular optical density units 
according to the following formula:
6.5.b Calibration Settings
After you have entered the step tablet values, you can immediately calibrate 
by clicking on the Calibrate Now button (in the Densitometer Options dialog 
box).
You can also specify how often you want the GS-710 to automatically 
recalibrate. Either click on the Calibrate Before Every Scan checkbox or enter a 
Recalibration Interval (in minutes) in the appropriate field.
Note: The scanner will automatically recalibrate each time you change your 
filter or your reflective/transmissive setting. (If you select a different 
application with the same filter and light settings, it will not auto 
recalibrate.)
Calibration Report
To print out a calibration report each time the densitometer calibrates, click on 
the Calibration Report checkbox.
6.6 Acquire the Image
Note: Before scanning in transmissive mode, make sure the white balance 
region of the scanning area is not covered or obstructed in any way.
To begin to scan, click on the Acquire button. The scanned image will begin to 
appear in the scanning window, line by line.
To interrupt a scan, click on the Stop button. A message will ask you to 
confirm the interrupt, and then you will be asked if you want to keep the 
partial scan. This feature is useful if you overestimated the size of the area 
you selected.
Specular OD 1.4 Diffuse OD⋅=