User guide
Fluor-S MultiImager
8-13
Referenced
If you do not want to perform a dark exposure with each acquisition, you can 
take a “reference” dark exposure that will be saved and subtracted from all 
subsequent acquisitions. Click on the Referenced button to activate this 
feature.
The first time you acquire an image after selecting this option, the Fluor-S will 
take a 60-second dark exposure that will be saved and used to subtract the 
dark count from all subsequent acquisitions.
For image exposures of greater or less than 60 seconds, the reference dark will 
be scaled accordingly and then subtracted. You can change the default 
reference dark exposure time using the Reset Reference button (see below).
If you deselect the Referenced button and then reselect it, the old reference 
dark exposure will still be available.
Note: Separate reference dark exposures will be taken for High Resolution 
mode and High Sensitivity mode. Once you have created a reference dark 
in each of these modes, each reference dark will be used according to the 
mode you are in.
Reset Reference
If you would like a reference dark with an exposure time that more closely 
matches that of your typical scans, click on the Reset Reference button.
Fig. 8-10. Reset Reference Dark pop-up box.