User guide
Chemi Doc
4-7
Fig. 4-5. Setting a manual exposure.
With Manual Expose activated, you can adjust the exposure time directly by 
changing the number of seconds in the Exposure Time field. Type in a number 
or use the arrow buttons next to the field.
In UV or White image mode, when the specified exposure time is reached, the 
last captured image will be displayed in the Chemi Doc image window. The 
camera will continue to integrate the image on the CCD, updating the display 
whenever the specified number of seconds is reached.
Once you are satisfied with the quality of the displayed image, click on the 
Freeze button to stop the exposure process. The last full exposure will be 
displayed in the image window.
Fig. 4-6. Freezing the manual exposure.
Manual Expose active;
button appears selected
Exposure Time field active;
adjust number of seconds
Click on Freeze to stop
the exposure process