User guide
GS-700 Imaging Densitometer
5-11
described below. You do not need to change the values in the reflective 
step tablet form; you can use the default values.
With Calibration On selected, the other calibration settings become active.
Calibration Strip Window
When calibration is turned on, a calibration strip window will appear below 
the main scanning window and the length of the main scanning window will 
be reduced to 29 cm reflective and 23 cm transmissive.
Every time the densitometer calibrates, an image of the calibration strip will 
appear in the calibration strip window.
5.6.a Editing the Step Tablet
Before you can calibrate, you must enter the step tablet values for your 
transmissive calibration strip into the Step Tablet Values form.
In the package with your calibration strip, you will find a printout of the 
diffuse density values for each segment of your transmissive calibration strip. 
Those exact density values must be entered into the computer for the 
software to associate a correct density value with each step on the step tablet.
Note: Scanning in transmissive mode with incorrect step tablet values entered 
into the form can cause significant errors in the reported densities of your 
scans.
To enter the density values for your step tablet, click on the Edit Step Tablet 
button in the Densitometer Options box. The Step Tablet Values dialog box for 
the GS-700 will open.










