FIGURE 12 – CHROMATOGRAPH OUTPUT EXAMPLE
Double left clicking on the gas chart will bring up a select tool that changes color and is “stuck” to the tip of the
cursor. Also at the tip of the cursor is a green information box that displays the time in minutes. If the cursor
goes to the left of the geolograph tick marks, the green information box displays the depth. This is useful in
selecting a range of feet or minutes of data. By double clicking on a second point on the Gas Chart, you have
just selected a range of data, and the Select Items Box will appear.
Examine Raw Chromatograph This
button allows the user to see a
close-up view of the selected
chromatograph output. Left
clicking and holding on the
chromatogram will allow the user to
adjust the trace up and down on
the display window. If the trace is
too large or too small, the user can
adjust the overall visual size in the
Program and Equipment Setup
Screen, under the Chromatograph
Tab. When the user is satisfied
with the display, a print can be
made of the chromatogram.
The peak search times are shown as heavy-colored overlays on the chromatograph curve. Each time-window
is centered on its associated vertical line.
The data for the particular chromatogram is available to the right of the chromatogram. If a particular peak
timing is off, or is incorrectly labeled, then all the user has to do is click on the Set Cx peak, place the cross
hairs over the correct peak, and DOUBLE CLICK on the peak to set the new timing window for that
component.
To see the previous separation, hit the Previous button. To see the next separation, hit the Next button.
Calibrating or adjusting the flow rate:
If you want to calibrate the flow rate, get a flow meter. The best one to calibrate with is a clear plastic
flowrater. It should be labeled 0 - 5 LPM. Omega Part number FL-2039.
From a console connection to the Bloodhound type in "calibrate flow" and enter.
This will put the Bloodhound into a calibration mode that will allow you to adjust the flow.
Rotate the black knob on the face of the Bloodhound clockwise to increase the flow rate. Adjust to 1LPM on
the plastic flowrator using the external flow meter.
Wait until the process times out. At that time the Bloodhound will internal electronic flow meter will have been
calibrated.
To just manually adjust the flow rate of the Bloodhound, set register 10 to a pump power number from 1 -
100. This will set the sample pump percent motor power level. The Bloodhound will not be able to
automatically adjust the flow rate when the flow rate is in this override mode.
Select Items Box