Specifications

Copyright 2014 Carl Bright Inc. All Rights Reserved
Page 29
DRILL DEPTH IS: 2813
To set the depth counter, enter the command followed by the new depth.
>Setdepth 2800[ENTER]
DRILL DEPTH IS: 2800
CGTEMP
This command will allow the user to set the column temperature remotely or through the console. To use just
enter the command with the desired temperature to run the column at. NOTE: Do not generally run the column
over 210 degrees F. There is a thermal fuse set at 240 degrees F. Temperature is in degrees F.
CGTEMP 150[ENTER]
CGSAMPLE
This is also a troubleshooting command. When issuing this command the Bloodhound will automatically
restart a chromatograph cycle no matter where it is in the chromatograph timing. This command is used to
help calibrate and troubleshoot the chromatograph.
CGASMPLE[ENTER]
WRITEETIME
This command is used to individually set the timing of the CG center point (peak), in seconds.
WRITEETIME C1 70
This example would set the peak timing for C1 to 70 seconds.
MUTE
The mute command will allow the remote user to silence or turn off the local alarm tone generator.
To set the mute for a number of minutes, just add the modifier.
MUTE 0[ENTER]
Will turn on the alarm tone if an alarm is present
MUTE 5[ENTER]
Will mute the Bloodhound for 5 minutes
MUTE 1000[ENTER]
Will turn off the Bloodhound alarm.
NOTE: If there is no present alarm, the tone will not be sounding, even if set to 0. Setting MUTE to a number
greater than 1000 has no effect.
?
Sending a question mark [enter] to the console will return the string Bloodhound and the model number 4.
This allows for the charting software or any other software to check to see if there is a Bloodhound connected
to that port.
?[ENTER]
Bloodhound 4