Specifications
Copyright 2014 Carl Bright Inc. All Rights Reserved
Page 34
REG. FUNCTION Units Range Factory
Setting
COMMENT
28 LO & HI gas detector
independence
0 OR 1 0 IR independence register : This register sets up
the way the high range and low range interact.
If this register is set to 0 then the IR ranges are
not independent and are joined This means
when the gas level reaches the end of the low
range it then goes to the high range IR sensor,
but it calculates out the high range sensor level
at the low level point. This attaches the high
point of the low range calibration to the low
point of the high range calibration to help
calculate the new response curve. If this
register is 1 then the IR sensors are
independent from each other. This means
when the gas level reaches the end of the low
range then it jumps to the high range where
ever it is at on the gas level concentration. This
usually shows up as abnormally large high
range gas units. The BH enters this mode and
cancels the dampening when in calibration
mode. Setting this register to 1 will basically
show the calibration points when the calibration
gas is injected.
29 Power down timer MIN 120 This is the number of minutes before power
down when the AC power is lost.
30 GAS READING
ATTENUATION
% 1-200 100 This is a percentage of the calibrated output - It
has the same effect as air dilution. 100 = 100%
output, 50 = 50% output, and so on.
32 Feet per geolograph
closer
FT 1 This is the number of feet incremented per
geolograph contact closure.
33 Geolograph noise spike
count alarm level
180 If there are noise spikes on the geolograph line
and the spike count is greater than this number
then the Bloodhound will alarm with a noisy
geolograph alarm. If set to 0 this alarm is
disabled.
34 Foot tick chirp milli-
seconds
0 Sets the time in milliseconds that the sonalert
will chirp when a foot is drilled. The larger the
number the longer and louder the chirp.
36 HIGH GAS RANGE
ATTENUATION
30 Same as register 30 but for the high range. Is
additive to register 30. Output = (adjusted low
range +
(corrected and adjusted high range x
(reg36/100) ) ) x (reg 30/100)
37 High Amps Alarm on
12V Cavitator terminals
Note:
60-sec
average
5 Set to 0 to disable alarm.
39 Low Amps Alarm 0-1 1 0 = off, 1 = activate alarm
41 H2S sensor Defeat 0-1 1 0 = no data - off, 1 = on (firmware >= 170).
Every time the Bloodhound power is cycled this
register is reset to 0 to default to the “off”
position.
If Register 41 = 255 the H2S sensor will remain
on after a power cycle of the Bloodhound.