Specifications
Copyright 2014 Carl Bright Inc. All Rights Reserved
Page 35
REG. FUNCTION Units Range Factory
Setting
COMMENT
42 O2 Sensor Defeat 0-1 1 If this register is set to 0 then the O2
sensor is effectively off and will read
zero. Also the LCD screen will show
OFF instead of a percent number and the
WITS packet 1225 will not be sent out
the WITS interface.
43 CO2 Sensor Defeat 0-1 1 If this register is set to 0 then the
CO2 sensor is effectively off and will
read zero, also the LCD screen will
show OFF instead of a percent number and
the WITS packet 1222 will not be sent
out the WITS interface
45 ALARM MUTE TIMER MIN 1-25 10 Mute timer in minutes. NOTE: not to exceed 25.
This is the number of seconds that the mute key
will mute the alarm tone.
46 High internal temp level 120 High internal temp level usually set to 120 deg F
47 low internal temp level 35 low internal temp level usually set to 35 deg F
50 High sample flow alarm
setting
LPH 200 Manual says "Normally set to 90"
51 Low sample flow alarm
setting
LPH 20 Manual says "Normally set to 40"
52 High Vacuum level
alarm setting
mmHg 200 200 means -200 mmHg
53 High gas units alarm
level
22 High gas units alarm level / 100. This is the gas
units alarm level divided by 100. So if this is set
to 10 then the alarm level is 1000 gas units.
Set to 0 to turn off. Factory set at 22 (2200 gas
units)
54 Chromatograph low
pressure alarm setting
mmHg 70 When the chromatograph pressure falls below
10 times this level an alarm is set. Set to 0 to
disable. (70 = 700 mmHg)
55 3G MODEM ON OR
OFF
1 1 - Internal modem is ON
0 - Internal modem is OFF
56 H2S alarm level in parts
per million / 10
PPM 0-200 50 0 = off. IF the H2S is above this level then the
alarm will sound. The alarm is off, not the
sensor. Go to Reg 41 to shut off the sensor.
This sensor is divided by 10, thus 50 = 500 ppm.
1 = 10 ppm alarm levels.
57 External alarm relay
control.
0 or 1 0 If this register = 0 then the alarm relay will close
on all alarms. If this register = 1 then the alarm
relay will close on just the H2S alarm. This does
not effect the normal alarms.
58 GC AVERAGING TO
ELIM NOISE FLOOR
1-35 1
59 debug messages
control
0 This will send debug messages over the given
port: 0 = debug messages off; 1 = PC PORT; 2
= WITS PORT; 3 = MODEM PORT; 4 = SB72
PORT; 5 = off; 6 = HRM PORT
61 C1 timing window
before and after peak
time
6 C1 timing window before and after peak time
(vert line)
62 C2 timing window
before and after peak
time
6 C2 timing window before and after peak time
(vert line)