Sucked Water Into
Instrument
If Pri and Sec Volts = 0.010, instrument has been flooded - return to iBall, or try:
Note: There is no guarantee this procedure will work, so start working on your
Plan B now. Turn off the instrument and disconnect the sample tubing from the
instrument. Remove 4 screws holding instrument cover and lift cover towards you.
Replace sample line filter located inside left front of instrument with a new filter.
Replace cover and power up instrument. Set Register 10 to 199. This runs the
sample pump at full power, which may help dry out the instrument’s internals. If you
have a drying tube (Drierite or equivalent), connect it to the sample intake port to pre-
dry the air entering the instrument. Write down the current Chromatograph Column
temperature and set the Chromatograph Column temperature to 190 degrees. This
will help dry out the column. After an hour running under these conditions, reset the
Column temperature to what it was before and set Register 10 to 0, which sets the
sample pump to automatic operation. Zero the Total Gas reading and test the
instrument with test gas or a butane lighter.