Technical data

- 61 -
P1.1 INT. CALIB. Internal adjustment
Start of internal adjustment process. Adjustment is conducted automatically,
and no user assistance is required. If weighing pan of the balance is loaded,
balance will display a command to unload the weighing pan. If weighing pan is
loaded with a relatively small mass (up to 4% of balance max capacity) then
adjustment process will be carried out automatically, but measuring results
may differ by the amount that was present on balance weighing pan.
or
P1.2 EXT. CALIB. External adjustment (with an external weight)
Adjustment with external weight, which value is saved in factory menu of
balance, function unavailable in verified balances.
CAUTION
Function unavailable in verified balances.
P1.3 USER CALIB. User adjustment (with an external weight)
Adjustment with external weight at any mass in the balance range, however,
not lower than 30% of max range.
CAUTION
Function unavailable in verified balances.