24. PIPETTE CALIBRATION
Balance allows pipette calibration by means of a special balance software or by
means of cooperation with computer software “Pipettes” intended for pipettes calibration
(workstation for pipettes calibration).
Before pipettes calibration a special set must be mounted inside the draft shield. The
set is an atypical balance equipment. See the picture below for set assembly instruction.
Balance with a set for pipettes calibration:
Calibration set assembly procedure:
dismantle the weighing pan and the draft shield,
place centering ring (8) inside the draft shield,
put glass ring (7) onto the centering ring,
place weighing pan (6) inside the glass ring,
place top ring (5) onto the glass ring,
place evaporation ring (4) onto the top ring (5)
insert calibration vessel (3) into the evaporation ring,
place the glass cover plate (2) onto the top ring,
put glass lid (1) onto the glass cover plate.
Balance with a mounted pipette calibration
set.
Evaporation ring eliminates measurement errors being a result of liquid evaporation
occurring while weighing process.