User manual
- 155 - 
27.13. A report on product control 
An example of a report: 
----- PCG Report: U/22/08/12/14/22 -----    
---------------------------------------- 
Balance type   PS 3Y   
Max capacity   1000 g    
Reading unit    0.001 g      
Balance ID   9908684     
Start date    2012.08.22 14:20:11    
End date   2012.08.22 14:22:35     
User 
Product    1     
Batch no.   555    
Nominal mas   10 g   
Tare  10.0002 g    
Value of T1 error   0.9 g    
Value of 2T1 error   1.8 g  
Value of +T1 error   g   
Value of +2T1 error    g    
Batch quantity    100    
No. of measurements  30   
No. of T1 errors   0    
No. of 2T1 errors   0    
Min   49.999 g   
Max  50.001 g   
Average   50.00057 g    
Total  1500.017 g   
Average limit 9.99969 g  
Average limit [+]  g 
Standard deviation 0.000626 g    
PGC mode  Non-destructive average tare 
Result  Positive 
Measurements 
1. 49.999 g      
2. 49.999 g      
. 
. 
28. 50.001 g     
29. 50.001 g      
30. 50.001 g      
----------------------------------------   
Signature 
........................................   
Report template: 
The balance user in the submenu < Printouts/  Printout template of a 
CPG report > can edit a template of a report on product control. The variables 
attributed to <YES> are printed in a report. 