User manual
- 197 - 
Example of a report: 
.....................................................    
--------- Autotest GLP: Report ---------
Balance type   PS 3
Y   
Balance ID 400010   
User   Admin    
Software rev. L0.0.21 S   
Date 2012.01.16   
Time 09:17:16   
----------------------------------------
Number of measurements   
10   
Reading unit   0.001/0.01 g   
Internal weight mass 1402.094 g   
Filter  Average   
Value release  
Fast & Reliable    
----------------------------------------
Deviation for Max.  -
0.118 g   
Repeatability 0.0088 g 
Signature       
........................................ 
PROCEDURE: 
On pressing a field with test name, the balance displays a message box. 
On this level, the user can carry out the following: 
  Start the following AUTOTESTU GLP process 
  Preview results of completed autotests, and export all saved test results as 
file with extension *.csv 
  Delete all saved test results 
User can abort the process at optional moment by pressing <X> key in the 
testing window. 
Results of carried out autotests are displayed as a table (each line features date 
and time of autotest procedure and its result).  
To preview detailed data of an autotest, press a specified line of the table with 
test results. 
In order to print a result of a single autotest, enter its details (see above) and 
generate a printout by pressing <print> key in the upper bar of the balance’s 
display.  
Autotest results can be exported by pressing <export> field on the level of all 
saved reports in a table. Data is sent to a data storage device connected to 
balance’s USB port as a file with extension *.csv.  