User manual
- 205 - 
Frame format 
1 
2 
3 
4 
5 
6 
7-15 
16 
17 
18 
19 
20 
21 
S  U  I  stability  space  sign  mass  space  unit  CR  LF 
  Example: 
  S U I CR LF – command from a computer 
  S U I ? _ - _ _ _ 5 8 . 2 3 7 _ k g _ CR LF - command carried out, 
  immediate response of mass value in current measuring unit 
  Where: _ - space 
  Switch on continuous transmission in basic measuring unit 
  Format: C1 CR LF 
  Accessible responses: 
C1_I CR LF 
- command understood but not accessible at this moment 
C1_A CR LF 
MASS FRAME 
- command understood and in progress 
- response is mass value in basic measuring unit 
  Frame format 
1 
2 
3 
4 
5 
6 
7-15 
16 
17 
18 
19 
20 
21 
S  I  space 
stability 
marker 
space  character  mass  space  unit  CR  LF 
  Switch off continuous transmission in basic measuring unit 
Format: C0 CR LF 
  Accessible responses: 
C0_I CR LF 
- command understood but not accessible at this moment 
C0_A CR LF 
- command understood and in progress 
  Switch on continuous transmission in current measuring unit 
  Format: CU1 CR LF 
  Accessible responses: 
CU1_I CR LF 
- command understood but not accessible at this moment 
CU1_A CR LF 
MASS FRAME 
- command understood and in progress  
- response is mass value in current measuring unit 
  Frame format: 
1 
2 
3 
4 
5 
6 
7-15 
16 
17 
18 
19 
20 
21 
S  U  I 
stability 
marker 
space  character  mass  space  unit  CR  LF 