User manual
- 45 - 
6.  Press <Method> field 
7.  Which opens a window containing a list of minimum sample methods 
stored in balance database. 
8.  Select one of available entries. 
9.  The software returns to the previous menu window 
10.  Press <Mode> field 
11.  Which opens a window with parameter settings. Select one of below 
options: 
Lock – on selecting this option and during the weighing process the 
display indicates corresponding pictograms that inform a user whether 
the weighed mass is below or above the minimum sample limit. 
Additionally the software disables accepting the measurement result if it 
is below the minimum sample limit 
Warn – on selecting this option and during the weighing process the 
display indicates corresponding pictograms that inform a user whether 
the weighed mass is below or above the minimum sample limit. The 
user can accept a measurement that is below the minimum sample 
limit, but on a printout such measurement is preceded by an asterisk 
(*). 
12.  On setting the parameters exit the menu. 
13.  The mass field on the display is supplemented by an additional 
informative pictogram. The pictogram changes during weighing process 
and indicates current status of mass in relation to the pre-determined 
value of minimum sample. 
 - Mass below the set value of minimum 
sample 
 - Mass above the set value of minimum 
sample for a given tare range 