User manual
- 48 - 
•  Acting according to the command, place an ordered weight/standard on 
balance’s weighing pan and press   key, 
•  On completing the adjustment procedure the balance indicates a message 
box: 
•  On confirming the message box with   key, the balance returns to 
weighing mode.  
11.3. User adjustment 
User adjustment is carried out with an optional standard, which mass ranges 
between 0,15 Max and Max. Adjustment procedure is compatible with the 
external adjustment process, but before its start the software opens a 
message box for entering mass of a standard used for user adjustment.  
  CAUTION: 
User adjustment is available only in balances which are not subject to conformity 
assessment (verification).  
In order to start user adjustment, enter submenu < Adjustment >, and select 
option: “   User adjustment”. Then follow the commands indicated on 
balance’s display. 
11.4. Adjustment test 
< Adjustment test > function enables comparing the result of internal automatic 
adjustment with the value of internal weight saved in balance’s factory 
parameters. The comparison is used for determining balance’s sensitivity drift 
over time.  
11.5. Automatic adjustment 
This menu option is used for selecting a factor which determines start of 
automatic adjustment process. Accessible options:  
  None – automatic adjustment disabled 