e)  Select option <Assign standard>, reference mass is saved in a product 
record under entry <Mass>. 
16.7. Part counting procedure 
The first step in parts counting mode is obtaining data on mass of a single 
part. Select one of available options: 
−  Give mass value of a single part (see point 16.3) and place parts on 
balance’s weighing pan, balance displays totalized parts.  
−  Set mass of a single part from a given quantity of parts (see point 
16.4.), the balance additionally shows pictogram of the ACAI function 
< > (if enabled). Place parts on balance’s weighing pan, the balance 
displays totalized parts. 
−  Acquire mass of a single part from database of products (see point 
16.5.) by selecting a desired product record. Place parts on balance’s 
weighing pan, the balance displays totalized parts. 
Caution: 
All additional elements (i.e. packaging) have to be tarred before starting the parts 
counting process. 
16.8. Checkweighing function in parts counting mode 
Parts counting process can be aided by checkweighing function, i.e. control 
whether indication is within set thresholds.   
  Checkweighing requires setting values of two thresholds, described as: 
−  LOW threshold [Min= … parts] 
−  HIGH threshold [Max= … parts] 
And enabling bargraph (set its attribute to YES), which displays below 
relation: 
CURRRENT NO. OF PARTS / CHECKWEIGHING THRESHOLDS. 
Defining the values of Low and High thresholds is carried out in the Database 
of Products while editing Product entry, or using a quick access key <  
Defining thresholds>. 