User manual
- 81 - 
3. If tolerance applies, set its value, 
4. Accept by pressing < > key, 
5. Press High Threshold key and insert its value in [%], 
6. Accept by pressing < > key 
19.3. Interpreting the function by use of a bargraph 
The dosing and checkweighing functions are aided by graphic indication, in a 
form of a bargraph. Below see an example of simultaneous operation of the 
two functions.  
a)  checkweighing thresholds< > are set as Low Threshold = 90%,  
High Threshold = 110% 
b)  Target value = 105% ; tolerance = 5% < > 
c)  Reference mass = 19.986 g < > 
19.4. Additional settings of Percent Setup mode 
The additional settings enable adjusting the working mode to user’s needs and 
requirements. Functions related to checkweighing: result control, tare mode, 
automatic footer printout, printout mode / value release, printouts.  
Means of using the above functions are given in point 10.8. “Additional 
parameters on weighing process”.  