12.4.2. External adjustment
External adjustment in balances APPC series should be carried out with an
external mass standard / weight class F
2
Start of external adjustment process,
Balances with internal adjustment balances with external adjustment
Balance will display a command ordering unloading of weighing pan unloAd
(weighing pan should be empty). When weighing pan is unloaded, press
PRINT key. The balance determines mass of empty pan. Place an external
adjustment weight with mass determined on balance’s display and press
PRINT key. On completing adjustment process the balance returns to
submenu.
Go back to weighing mode with procedure of saving changes
(see point - 8. 2. – return to weighing).
If a balance is verified, then user does not access to carrying out
external adjustment process.
12.5. Adjustment report printout
On completion of any type of adjustment process, the balance enables
preparing a report from adjustment process. The report can be printed on a
connected printer and sent to a computer and saved in a form of file for
records. Printout process is described in details in point 9.2 of this user manual.
The user can select between two options.