User manual

- 81 -
16.2. Printing date with specified date and time
Every weighing result can be printed with specification of measurement date
and time. The presence of date and time on a printout is conditioned by
connection of printers type Kafka 1/Z or Kafka SQ S to a balance.
When connected, go to balance menu and set parameters from group P2 GLP:
PdAt - YES
Ptin - YES
16.3. Cooperation with statistic printer Kafka SQS
On connecting a statistic printer KAFKA SQ S to a balance, it is possible to
carry out statistics from completed measurements. An instance of a printout
including statistics from series of measurement:
1 9:02:15 + 7.0016 g
2 9:02:39 + 5.0152 g
3 9:02:58 +
12.0171 g
4 9:03:15 + 9.9937 g
5 9:03:34 + 12.0169 g
6 9:03:48 + 22.0111 g
Date 13.09.2001 Time 9:04
n 6 number of samples
sum x 68.0556 g total mass of samples
x
11.34260 g average value
s 5.92328 g standard deviation
srel 52.22 % variance factor
min 5.0152 g min value
max 22.0111 g max value
R 16.9959 g max min difference
16.4. Format of data sending
Weighing result can be sent to a peripheral device by pressing PRINT key on a
balance keyboard or by setting a control command from computer level.