User manual no.: LMI-61-05/04/13/ENG Balances 3Y series • Microbalances – MYA.3Y BALANCES AND SCALES RADWAG 26 – 600 Radom, Bracka 28, POLAND Phone: +48 (0-48) 38 48 800, fax. +48 (0-48) 385 00 10 export@radwag.com www.radwag.
APRIL 2013 -2-
TABLE OF CONTENTS 1. BASIC INFORMATION ............................................................................................. 6 1.1. Intended use .................................................................................................................. 6 1.2. Precaution measures ..................................................................................................... 6 1.3. Warranty ..................................................................................................
11. DISPLAY CALIBRATION ..................................................................................... 32 12. BALANCE USER SELECTION ............................................................................. 32 13. PERIPHERALS ..................................................................................................... 33 14. ABOUT .................................................................................................................. 33 15. COMMUNICATION ........................
18.16. Mass Comparison .................................................................................................... 95 18.16.1. Select method for comparing mass ............................................................................................. 96 18.16.2. Mass comparison - measurements ............................................................................................. 97 18.16.3. Mass comparison – report from a measuring series ..................................................
1. BASIC INFORMATION 1.1. Intended use Balances 3Y series are intended to precise mass measurement of weighed loads conducted in laboratory conditions. It is designed only for non-automatic weighing processes, thus the material to be weighed needs to be placed on the center of balance’s weighing pan with required carefulness. 1.2. Precaution measures Never use the balance 3Y series is areas endangered with explosion. Balance 3Y series is not designed to operate in EX zones.
2. TRANSPORT AND STORAGE 2.1. Packaging Retain all elements of balance packaging in case they are needed for future balance transporting. Only an original balance packaging can be used for transporting a balance. Before packaging, disconnect all cables and conductors from the balance and remove any moveable parts from the weighing chamber (weighing pan, shields, inserts). Balance components should be packed into their original packaging for their proper protecting against possible damage during transport.
Working temperature: 10-40 °C Relative air humidity: 45% - 65% Because of the structure of the objects to be weighed, it may be necessary to store them in glass containers for eliminating the phenomenon of humidity absorption.
3.2. Unpacking Carefully remove the balance from its packaging, remove the plastic and foil transport protective elements. Gently place the balance in its intended place of use. Install the weighing pan, and other elements according to below scheme. Microbalance MYA 3Y series After installing the balance in its intended place of use assemble its components following above scheme and in relation to balance type. A – standard 1. 2. 3. 4.
After the elements setting, connect any additional devices and only then plug the balance to power source. Power socket is located on back side of balance terminal containing the display. Balance for pipette calibration: After balance start up change the standard weighing pan into the one for placing a glass vessel (5) and on stabilization of mass indication press TARE button. 3.3. Balance cleaning Clean the balance using a damp ch by gentle rubbing any contaminated places.
4. GETTING STARTED Switching on the balance: Press «On/Off» button on terminal overlay. The balance carries out an internal testing procedure, and after its completing it is ready for use. Caution: If the balance is incorrectly leveled, then on switching on it displays a alert ordering leveling the balance. The leveling procedure on described in ch. 5.
5.1. Level control system – automatic operation – “ALERT” mode If the level sensor detects any level change, then the colour of the level pictogram on the display changes into red (colour of the smallest dot in level pictogram), and an alert message is displayed. In order to level the balance press “ENTER” button in the displayed message box.
5.2. Manual leveling If the electronic level control function is disabled in balance menu, then in case of level change a balance user has to manually enable the leveling procedure. Press pictogram of the level status. The display changes into a control panel of the leveling function with a top view of current level status visible on its right side, and balance elevation on the left.
6. MENU STRUCTURE 6.1. Working mode selection Each balance operator can set their individual working mode enabled as DEFAULT on activating their profile. As a result, the balance starts operation with a working mode specific for an enabled user profile, i.e. in parts counting mode or percent setup.
6.2. Moving in user menu A balance 3Y series features a colourful touch screen display making balance software operation very simple and intuitive. or Enter main menu Quick scrolling menu to the top Scroll menu upwards Scroll menu downwards, Quick scrolling menu to the bottom Accept changes Leave a function unchanged (without saving changes) Add item to a database (e.g. database of operators) Delete item from a database 6.3.
6.3. Quick return to weighing mode The software enables quick returning from an optional menu section with or without saving made changes. Return to menu WITHOUT saving changes. Press and hold button for app. 2 sec. located in the upper right corner of the display. The software returns to mass displaying window. Return to menu WITH saving changes.
User menu structure comprises 8 main groups accessed by pressing SETUP button on terminal overlay or Adjustment pictogram on the touch screen display.
− User parameters user password language administrator print settings copy settings − Menu look colour screen brightness additional display brightness beep sound − Readout filter value release refreshment autozero last digit − Keys and sensors left proximity sensor right proximity sensor sensor sensitivity sensor sound keyboard sound effect − Date/Time date format time format time date show time show date − Printout standard printout ο header ο line ο fo
7. WEIGHING Simple weighing process is carried out using only the buttons located on the right hand side of the balance terminal. The balance features separate buttons for zeroing, tarring and generating a printout. 7.1. Zeroing process Pressing the zeroing button causes determining a new zero point for a balance. From now on mass of all loads placed on balance weighing pan (including tare) is determined from this point. After zeroing the displayed indication, the tare and net (gross) values equals 0.
7.3. Use of constant tare in weighing If a weighing process applies a set of the same packages with determined mass then the balance enables determining a constant tare. On weighing process the constant tare value is recalled by pressing a dedicated QUICK ACCESS KEY. The value of indicated measurement result is net, i.e. displayed with already applied tare.
o Press ADD pictogram to add a new tare item o Enter name and value of a tare o Changing tare value is carried out by pressing field with displayed tare value which opens an on-screen numeric keyboard o Enter the tare value and accept it by pressing OK. o Changing the tare name is carried out by pressing a field with tare name which opens a on-screen keyboard for entering name of a tare. Tare name may contain up to 15 alphanumeric characters.
7.4. Applying automatic tare in weighing Means of operation: Each first measurement is tarred immediately after stabilization, that is signaled by NET pictogram appearing on the display. The balance indicates net mass of weighed load.
8. BASICS ON TERMINAL AND SOFTWARE OEPRATION This chapter covers details on terminal display and means of terminal operation. The presented descriptions refer to all balance featuring this type of terminal independently on their metrological parameters. TERMINAL DESCRIPTION 1. Infrared proximity sensors The terminal features two proximity sensors. Both of them are optionally configurable to operate as buttons or menu functions (e.g. opening weighing chamber doors, zeroing, printing, etc.).
4. «F3» button Pressing the F3 button triggers automatic internal adjustment function. 5. «SETUP» button Pressing the SETUP button enables determining main software settings referring to all applications operated by a user. 6. «On/OFF» button The ON/OFF button enables turning on and off balance display, thus switching to the stand-by mode. 7. «0» button Button for zeroing mass indication on balance display. 8.
3. date / time Indicates current date and time. Pressing this display field opens a menu for changing date and time settings. 4. level status A level pictogram informing on current level status of the balance. Pressing this display field opens a leveling assistant. Description of the leveling process is provided further in this user manual. CAUTION: Flashing dot of the level pictogram indicates warming-up process of the level indicating electronic components. Warming up time is approximately 2 hours. 5.
8.1. Basics on balance operation 1. Each balance user can adapt the balance settings to the individual needs, including functions, printouts, colours, measuring units, etc. 2. Each working mode features its DEFAULT settings related to the QUICK ACCESS keys 3. Each working mode features individual settings of the workspace area for displaying data on a weighing process. 4. Each user can adapt a balance to display individual QUICK ACCESS keys and custom data in the workspace area.
9. BALANCE SOFTWARE The software of a balance 3Y series features the following components: User profiles System settings Working modes 9.1. Operators / User profiles Balance 3Y series features user profiles for adapting the balance functions and applications to individual user needs and specific tasks. A user profile is a set of customized settings that are saved in balance memory and recalled by a user by pressing a single button.
9.3. Working modes Working modes are software modules designed to carrying out a determined task related to weighing process. A balance 3Y series comprises a set of versatile application designed by the manufacturer (e.g. regular weighing, calculating statistics, formula making process, percent setup, parts counting or density determining process). On turning on the balance the last active user profile is automatically enabled, together with last used working mode.
10. BALANCE ADJUSTMENT Ensuring the highest measuring accuracy of a balance requires periodical entering to balance memory a coefficient correcting balance indications in relation to a mass standard. The described process is known as balance adjustment.
1 Internal adjustment Pressing this pictogram initiates internal adjustment process. The process is carried out fully automatically, without operator intervention. If the weighing pan of a balance is loaded, then the software displays a command to take the load off the weighing pan. 2 External adjustment Pressing the pictogram causes initiating the adjustment process using an external weight. The value of the external weight is stored in balance factory memory.
10.1. Adjustment report settings A report from balance adjustment is a procedure for objective determining that a balance is functioning correctly, and thus a carried out measuring series demonstrates actual mass of a tested object.
11. DISPLAY CALIBRATION A balance 3Y series comprises a terminal featuring a touch screen display. The display operates my means of reactions for resistance changes, which is a reaction to touching a specific display field. As some software functions and settings require precise touching of a specific display field, the balance menu contains an option enabling calibrating the display.
13. PERIPHERALS Menu “Peripherals“ enables determining types of devices to be connected to a balance and configuring ports settings for connecting any peripheral devices. A selected printer and a communication port must comply with actual status of the connected devices. The available USB ports operate as standard configured IT devices and do not require any modifications of balance parameters. The settings available RS 232 interfaces are modifiable, i.e.
15. COMMUNICATION Menu “Communication” contains balance settings referring to cases where the balance is one of IT network components. A user should set means of balance addressing (i.e. balance address), IP address, subnet mask and gate address. The above data should comply with the settings of IT network to which the balance is connected. 16. UPDATE Balance update is in other words changing the software of a balance user – it does not change metrological parameters of a balance.
18. BALANCE ADAPTATION – USER SETTINGS This chapter covers the means on determining INDIVIDUAL settings for each of balance users. This functions enables optimum adapting the balance to user needs and specific tasks. The settings are saved and stored in an active user profile and they refer to operation of all working modes covered by the modified user profile. On enabling a user profile, the saved settings are automatically activated.
18.1. Adding a new operator Press pictogram “PLUS” located in the top bar of the display which opens a window for entering name of an operator. After entering the name, accept it by pressing OK button. The name of a newly created user is displayed in the panel with balance users. Caution: Adding a new balance user is not allowed for each balance user, and its accessibility depends on authorization level assigned to a user. 18.2.
Each balance user has assigned the following functions: User details Look Readout Operation of buttons Printouts Units Working modes Access level Caution: Each balance user may have some limitations in balance operation. The range of limitation is determined by the balance administrator – i.e. DEFAULT USER. Name of a selected user is visible in the upper bar of the display. 18.4.1.
Print settings This function refers only to an active user and it has the following format 18.4.2. Look ------------ User details ----------User Default Language Engligh Administrator yes This menu contains data related to the LOOK and functionality of balance’s display, such as: Colour The field displays a list of colour palette available for a user. The selected colour refers to the top bar on the display.
18.4.3. Readout Menu readout contains data related to WEIGHING PROCESS, such as: Filter Each measuring signal is subject to electronic processing before it is displayed. The filtering is carried out to achieving correct measurement parameters, specific to a stable signal, i.e. one that is ready for reading by a balance user. Balance user is able, in a limited range, to change the signal processing parameters by selecting one of available FILTER settings.
Display refreshment This function determines the amount of instantaneous measuring data to be displayed during a weighing cycle. The balance enables previewing all of instantaneous measuring data, i.e. then multiple transient states are previewed. Another extreme setting is previewing the lowest number of instantaneous measuring data. In such case, approximately three transient states are previewed on the display before reaching a stable measurement status.
Right proximity sensor Select one of available processes to be carried out by the proximity sensor, including: o disabled o printout o tare o open door /option available in a balance MYA.3Y series/ o close door /option available in a balance MYA.3Y series/ o change door /option available in a balance MYA.3Y series/ o units Sensor sensitivity The parameter is adjustable within the range from 0 to 9, and determines the distance for reaction of a proximity sensor. As standard, the value is set to 5– 7.
Time /set time / Press active display field to set time in format visible in the top bar of the display. After setting time, accept it by pressing OK button. For instance 11:23:32 >> requires entering time as: 11.23.32 and pressing button. Caution: Incorrect time values shall be ignored by the software, remaining the time unchanged. Date Press active display field to set date in format visible in the top bar of the display. After setting date, accept it by pressing OK button.
PROCEDURE: 1. Press field with name of a template to be edited (Header – Line – Footer) and select variables to be printed 2. If a printout is a non-standard one, create it.
BASIC PRINCIPLES OF USING PRINTOUTS 1. Press PRINT key on balance’s overlay to print variables that are contained in section LINE of the standard printout template, and if they are attributed = YES (see above list of variables in printouts). 2. Variables attributed as YES present in the HEADER and FOOTER are printed ONLY on pressing Print Header or Print Footer soft keys located in the bottom bar of the display.
Pictogram A – Print data contained in the header Pictogram B – Print data contained in the footer CAUTION: Each balance user can have their individual set of QUICK ACCESS KEYS enabled. 18.4.6.1.
18.4.7. Printouts – Non-standard printout Non-standard printout A non-standard printout comprises of two sections: TEXTS and VARIABLES First design an optional number of texts, and then form a non-standard printut template using the designed texts. PROCEDURE: 1. Press field named TEXTS 2. The display opens a window for adding texts 3. Press “ADD” field and add a required number of texts used for designing a non-standard printout template.
1. 2. 3. 4. An entered and accepted name of a text is displayed in the main window of printout settings. This means of operation facilitates printout designing process. A user works on an actual template containing a set of texts. 5. Principles for working with texts 1. highlighting a text Press and release a field with text name. The text is automatically highlighted by a frame. Now it is possible to add a new text before or after the present one. 2.
Entering content of texts - Press name/number of a text to which a content should be added - press pictogram “ADD” and select a position of a text line: NEW, ADD AFTER, ADD BEFORE. Select an appropriate option. An option add BEFORE or add AFTER requires determining a corresponding line for carrying out the process. Selecting a line is made by highlighting its field. A new line will be added before or after the selected line.
- enter a text /max 40 alphanumeric characters/ using the on-screen keyboard, and press ENTER button after completing the process. Caution: A button for enabling national characters is a multifunctional one – see below diagram.
Creating a non-standard printout Balance 3Y series enables creating up to 5 non-standard printouts. Creating a printout template is carried out by adding individual texts to a printout template. PROCEDURE - press “PRINTOUTS” field - press “ADD” pictogram - the display opens a window with a printout number - enter a custom printout name /number and accept it - a printout template is created. Caution: Each following printout template has its individual order number, e.g.
18.4.9. Working Modes – general information Function “Working modes” contains a list of all working modes available in a balance. On displaying a window with a list of working modes, a user can: − Select a default working mode − Determine conditions for using a working mode.
18.5.
18.6. Parameter of working mode WEIGHING – quick access keys Buttons to be determined in working mode WEIGHING. If a button attribute is set to YES then it is automatically displayed in the bottom bar of display below the section indicating the weighing result. It functions as a “HOT-KEY” i.e. a quick access key for the most frequently used processes. YES ADJUSTMENT “Adjustment“ : enabling adjustment using an internal adjustment weight.
Caution: Maximum number of quick access keys to be added in the bottom bar is 5. If more than five quick access keys are enabled then only the first 5 are displayed. 18.7. Parameter of working mode WEIGHING – information Information items are displayed in the grey coloured workspace area. The workspace enables containing up to 6 pieces of information. If more than six parameters are enabled, then only first 6 are displayed.
Weighing mode provides settings a set of additional options Automatic printout ο Enabled: YES / NO ο Minimum mass It is a mass threshold required for repeated enabling the automatic printout option. The mass indication has to drop below the minimum mass threshold to repeatedly enable the automatic printout function. Unit definition no. 1 − Enabled: YES / NO − process, i.e.
Print button Enables determining operating mode of «PRINT» button. Available settings: ο “Stable“ Pressing the « PRINT » button causes printing of mass indication only after it is stabilized. ο “Always“ Pressing the « PRINT » button causes immediate printing of mass indication, independently on the stability status. ο “Off“: The « PRINT » button is disabled. ID definition Menu “ID definitions“ enables activating up to 12 codes that are accessible on pressing a hot-key.
18.8. Parts Counting Parts counting mode is carried out in two stages. First, a user should determine /or acquire from balance memory/ the mass of a single part, and secondly use the mass of a single part to determine a number of counted parts placed on balance’s weighing pan. The PARTS COUNTING mode has some individual ACTIVITIES assigned.
Below description contains only those quick access buttons that are not present in working mode: WEIGHING. On enabling the “Parts Counting mode” the display indicates default set of pictograms for immediate initiating work, i.e.: − Set tare − Enable tare − Determine part mass /1 part/ − Determine standard quantity − Database of standards Set part mass Pressing this hot-key opens a window with on-screen keyboard for entering mass of a single part.
Give standard quantity for standard procedure Select standard quantity from a displayed message box or enter other value. The order for carrying out the processes is indicated on the display by corresponding messages. A user can select among the following options: − Optional − 10 parts − 20 parts − 50 parts − 100 parts Caution: When setting a standard quantity maintain the following condition: mass of 1 part > balance reading interval.
o Which opens a window for entering name of a part to be saved o Accept entered name by pressing ENTER button – a part is automatically added to the DATABASE of STANDARDS. Mass of the saved part is previewed next to the name of a saved standard.
18.8.2. Creating database of standards Use of the quick access key “Database of Standards” located in the bottom bar enables only adding standards to the database. It is not possible to delete any records from the database. Full access to the Database is possible from the internal menu of the balance and it is dependent on authorization level of a user.
18.8.3. ACAI function – Automatic Accuracy Correction The ACAI function is to automatically correct mass of a single part during parts counting process. it is aimed to increase the accuracy of counting process – mass of a single part is calculated from a higher quantity of samples, ensuring higher accuracy. First, enable the ACAI function.
18.8.4. Parameters of parts counting mode – information Information items related to weighing are displayed in the grey coloured workspace area. The workspace enables containing up to 6 pieces of information. If more than six parameters are enabled, then only first 6 are displayed. YES Bargraph A bargraph refers to operation of the filling and checkweighing modes in the parts counting mode.. The bargraph is a graphic interpretation of modes operation.
NO Non-standard data Option YES – information is indicated in the workspace area, Option NO – information is not present in the workspace area. Non-standard data project Defining a non-standard information item requires creating an optional text to be displayed in the grey coloured workspace area. Procedure: − press field on “Non-standard data project” − Press “ADD” pictogram − Press a text field to edit (create) a text − Enter an optional text and accept it by pressing ENTER button Note: 1.
18.8.5. Graphic interpretation of Dosing and Checkweighing modes CHECKWEIGHING A criterion for enabling a bargraph is setting: − A quick access key for “Checkweighing” mode − Information item “Bargraph” in the workspace area to YES. On enabling the above two parameters, a bargraph is added to the display and it is located below the measurement result. Colour of the bargraph indicates the current quantity of counted parts.
DOSING A criterion for enabling a bargraph is setting: − A quick access key for “Dosing” mode − Information item “Bargraph” in the workspace area to YES. On enabling the above two parameters, a bargraph is added to the display and it is located below the measurement result. Colour of the bargraph indicates the current number of parts in relation to the target value. For enabling the filling mode, press a quick access key and specify the target number of parts. The entered value must be a total one, e.g.
18.9. Checkweighing Checkweighing is a process for controlling mass of a load. The load has low and high thresholds determined, these are the checkweighing thresholds, marked as LO and HI. Quick access keys displayed in the bottom bar relating to the checkweighing mode. Set their attribute to Yes (present in the bottom bar) or No (absent).
18.9.1. Creating database of checkweighing thresholds The database of checkweighing thresholds contains numeric values for the thresholds and their optional names. The database enables adding or deleting the records if they are not used. Full access to the Database is possible from the internal menu of the balance and it is dependent on authorization level of a user.
18.9.2. Use of checkweighing thresholds When in weighing mode, it is possible to use the checkweighing thresholds saved in the Database of Thresholds or define custom thresholds. In case of using the Database of Thresholds, it is possible to: − acquire thresholds to be used − saving custom thresholds in the Database of Thresholds Caution: Using Quick Access Keys does not enable deleting thresholds records from the database.
PROCEDURE: o press Quick access key o to open a screen with data saved in the Database of Checkweighing Thresholds o search and select a required threshold and press field with its name, o the values of the checkweighing threshold are automatically enabled for use in the checkweighing mode.
18.9.3. Bargraph in checkweighing mode. A bargraph is one of the elements of the workspace area. A criterion for enabling a bargraph is setting its parameter to YES when declaring information items for the checkweighing mode. The acceptance field on the bargraph depends on the current settings of the low and high checkweighing thresholds. Bargraph colours: − yellow: mass below the low threshold − green: mass within the set thresholds − red: mass above the high threshold 18.10.
Other quick access keys function in accordance with the description provided in chapter 18.7. Parameter of working mode WEIGHING – information. Information items related to dosing: o bargraph o supplementary unit o tare o gross o counter o target value o statistics /difference, amount, total sum, standard deviation, etc./ o ID codes o non-standard data o non-standard data project Each information item features option YES or NO determining its displaying criterion.
o If the database does not contain any records press “Add” pictogram o the display opens a window with on-screen numeric keyboard for entering VALUE of TARGET MASS. Accept the entered value by pressing OK button. o After specifying the VALUE the records gets a default name to be changed by a user. o Press field with record name o Use alphanumeric keyboard to enter name of the record and accept it by pressing ENTER button.
Temporary custom target mass values PROCEDURE: o press Quick access key o which opens a screen with a field for entering the value of target mass o press field with the value of target mass and using a numeric keyboard enter the its value. Caution: The values are rounded to the value of balance reading interval. Numeric keyboard – description of buttons: - DEL: delete last digit - C: delete all digits - OK.
Use of Database of Target Mass Values PROCEDURE: o press Quick access key o to open a screen with data saved in the Database of Target Mass Values. o search and select a required target mass record press field with its name, o the value of the target mass is automatically enabled for use in the dosing mode. o the display indicates value of the target mass preceded with MINUS sign o the balance is ready to work with enabled target mass value .
18.11. Percent setup Percent setup is a mode intended to compare weighed load with a pre-determined reference mass. The process is expressed in [%]. When declaring the mass to be compared, a user should specify mass value, e.g. 50.0g, to be recognized by the balance as e.g. 100%. This is the reference point for the consecutive measurements. Additionally, the percent setup mode can have the DOSING and CHECKWEIGHING modes enabled.
o o o o o o o low threshold high threshold target mass statistics /difference, amount, total sum, standard deviation, etc./ ID codes non-standard data non-standard data project Each information item features option YES or NO determining its displaying criterion. CAUTION: Defining non-standard information items is described in chapter 18.8.4. Below description refers only to those buttons /pictograms/ that are not available in working mode “Weighing”.
In practice, it is a tolerance for the mass to be weighed, i.e. the acceptance field in which the dosed mass is recognized as correct. The tolerance thresholds may be asymmetric, e.g. − high threshold: +5% − low threshold: - 2% the values of the checkweighing thresholds will be displayed in the workspace area. Dosing Enter the value of the target mass to be reached during comparing the following loads with the reference mass (reference standard).
Database of reference mass The database contains list of reference mass values (standard) with determined: o mass values in grams that are the reference standards 100% o names for database entries In the main window of the percent setup mode, a user can: o acquire reference standards o add reference mass (standards) to be database ACQUIRING REFERENCE MASS FROM THE DATABASE: o press name of a database entry to be used in percent setup mode o the record is automatically activated and the balance returns to we
Full access to the Database is possible from the internal menu of the balance and it is dependent on authorization level of a user.
18.11.2. Use of reference mass (standard) in weighing. When carrying out a weighing process, it is possible to use the reference mass records saved in the Database of Reference Mass (Standards) or define temporary custom reference mass values.
18.11.3. Use of reference mass values in weighing A bargraph is one of the elements of the workspace area. There are two criteria for enabling the bargraph: setting “bargraph” parameter to YES when declaring information items for the percent setup mode enabling at least one additional mode “Checkweighing” or “Dosing” in “BUTTONS” menu. After enabling at least one of the modes, it is available in a form of quick access key on the bottom bar of the display.
5. Mass of weighed load was increased to 53,2g that is 106,4% of the reference mass value. The display status for this mass value is mass excess (bargraph colour is red). 6. – above the high threshold value. Caution; the tolerance of the high threshold is + 5% to the reference mass and the display indicates + 6,6% thus the indication IS OUT OF the acceptance field. 18.12. Animal weighing Animal weighing mode is intended to determining mass of an object that is moving on the weighing pan of a balance.
Information items related to animal weighing mode: o bargraph o supplementary unit o tare o gross o counter o filter o threshold o statistics /difference, amount, total sum, standard deviation, etc./ o ID codes o non-standard data o non-standard data project Each information item features option YES or NO determining its displaying criterion. CAUTION: Defining non-standard information items is described in chapter 18.8.4. 18.12.1.
The following measurements are started automatically on placing a load on balance weighing pan. The following criterion has to be met: after taking the load off the weighing pan, the indication has to fall below the set threshold value. When selecting the threshold value the user determines the accuracy required for measuring mass of an object (the measurement does not start from the point of the precise zero). Threshold value = 50 intervals. 18.12.2. Animal weighing mode – manual mode of operation 1.
18.13. Density of solids, liquid and air The Quick access keys displayed in the bottom bar relating to the density mode. Set their attribute to Yes (present in the bottom bar) or No (absent). Adjustment ID definition Set tare Solids Enable tare Air Print header Liquids Print footer Statistics Clear counter xxx Non-standard buttons Principle of use and function of the quick access keys has been described in chapter 18.6.
o non-standard data o non-standard data project Each information item features option YES or NO determining its displaying criterion. CAUTION: Defining non-standard information items is described in chapter 18.8.4. Principles for using other buttons related to the DENSITY mode are provided further in this user manual. Caution: Detailed description of procedures on determining density of solids and liquids is provided in the USER MANUAL of the density determining kit. 18.13.1.
o the selected density mode is automatically enabled on returning to weighing. Caution: after changing the working mode the default working mode is automatically changed /automatic change of the Density field value: e.g. liquids → solids/ 18.13.4. Density of air Density of air is one of parameters required for calculating a correction value for a measuring result related to air BUOYANCY. Another parameter needed to calculate the correction value is density of tested sample. 1.
18.14. Formulations The formulation mode enables designing up to 10 mixtures, each containing maximum 20 ingredients. Each formulation ingredient has its specific name and value that is entered to a formulation description. First a user needs to create a formulation and then activate it to use. Creating formulations is carried out from the internal menu of the formulation mode. The formula making procedure is described further in this user manual.
18.14.1. Designing a formulation Database of formulations comprises names of formulations and names of ingredients forming the formulations. When designing a formulation, first specify its name and then add ingredients to a created formulation record. As each ingredient record requires specifying its name and mass, the operator should know the complete content of the formulation.
o Mass of each ingredient is rounded up or down depending on the reading interval of a balance. o A designed formulation is ready to use on returning to main window of the Formulations mode o Return to weighing – press “RETURN” arrow or press field with working mode pictogram. Prompts Enabling prompts (setting their attribute to YES) causes displaying name of the following ingredients to be weighed. It facilitates the making a formulation that consists of multiple ingredients.
A example of formula making process PROCEDURE: o press Quick access key o which opens a screen with a list of saved mixtures o select a mixture to be prepared on a balance Caution: Mixture content can be previewed by pressing field with its name o selecting a formulation causes automatic returning to weighing and displaying data on a formula making process in the workspace area.
o o o o WORKSPACE AREA shows a message: Weighing finished. Continue The display shows total mass of prepared formulation. Preparing the same formulation can be repeated by pressing OK button. Preparing another formulation requires entering the database by pressing Quick access key DATABASE OF FORMULATIONS Use list of saved mixtures and select one to be prepared Caution: The WORKSPACE AREA may show SUM OF INGREDIENTS if this information item has been enabled for displaying. 18.15.
Information items related to statistics: o supplementary unit o tare o gross o counter o statistics /difference, amount, total sum, standard deviation, etc./ o ID codes o non-standard data o non-standard data project Each information item features option YES or NO determining its displaying criterion. CAUTION: Defining non-standard information items is described in chapter 18.8.4. 18.15.1. Internal menu of Statistics mode The internal menu comprises two options: DELETE and RESULTS.
18.16. Mass Comparison Mass comparison mode enables comparing mass of a standard with mass of a tested object. The mode applies two testing methods: − ABA [standard – test object – standard] − ABBA [standard – test object – test object – standard] Each method requires carrying out a measuring series. Number of measurements in a measuring series is determined from 1 to 20. In practice, the most frequently used number of repetitions is 6.
18.16.1. Select method for comparing mass The mode enables selecting between two testing methods: ABA and ABBA. Select name of a required testing procedure. PROCEDURE: o in an optional moement press grey coloured workspace area o to open a screen with internal menu of the MASS COMPARISON mode o scroll down to Measuring Series field o press Measuring Series field o Select a comparing method to be used o Select number of method repetitions. The available number of repetitions ranges from 1 to 20, e.g.
18.16.2. Mass comparison - measurements When carrying out mass comparison process follow commands displayed in the COMMAND BAR of the workspace area. The mass comparison mode enables determining standard deviation that is calculated on basis of measuring series ABBA or ABA, where : A – is a mass standard for control B – is a tested mass. The results are calculatded by the software in accordance with below tables and equations. No. 1 2 3 4 5 ….. n For the ABBA series B B A No. 1 2 3 4 5 …..
18.16.3. Mass comparison – report from a measuring series Individual measurements in a series and a report from a measuring series is not printed in an automatic manner. The criterion for printing of the information is set in a STADNARD printout – FOOTER. PROCEDURE: o Press field with name of a user.
Printout example: No. of measurements A B A Difference Average difference No. of measurements A B A Difference Average difference No. of measurements A B A Difference Average difference No. of measurements A B A Difference Average difference No. of measurements A B A Difference Average difference No. of measurements A B A Difference Average difference Standard deviation 1 from 6 500.001 g 500.045 g 500.001 g 0.044 g 0.044 g 2 from 6 500.001 g 500.046 g 500.001 g 0.045 g 0.045 g 3 from 6 500.001 g 500.
18.17. Differential weighing Differential weighing mode enables analyzing changes of mass in a single sample or multiple samples. The process is carried out by determining sample’s initial mass, and then the sample is subject to different processes, which outcome is separating or adding of some of sample’s ingredients to its initial status. Finally, the sample is repeatedly weighed (differential weighing).
The Quick access keys displayed in the bottom bar relating to the differential weighing mode. Set their attribute to Yes (present in the bottom bar) or No (absent).
o non-standard data project Each information item features option YES or NO determining its displaying criterion. CAUTION: Defining non-standard information items is described in chapter 18.8.4. Principles for using other buttons related to the DIFFERENTIAL WEIGHING mode are provided further in this user manual. Below list provides the functions of only those quick access keys, that are absent in working mode “Weighing”: Weighing A Start sampling of initial mass of the sample.
18.17.1. Defining a series Each series comprises of a name and samples. Data on each of the sample contain complete parameters on a sample, such as its current status, and already carried out process parts. The samples are added during the differential weighing process, and before starting the weighing process , or . Adding a stable takes place by entering its name and carrying out a selected weighing procedure.
o Use available options to select item o An opened window contains a list of saved series o If the database does not contain any records press “+” pictogram o To open a screen with keyboard for entering name of a new series. Accept the name by pressing button. o The software adds the new entry to the list.
18.17.2. An example of carrying out a differential weighing process On activating DIFFERENTIAL WEIGHING mode: • Select information items to be displayed in the grey coloured workspace area • Select appropriate quick access keys • Add a to balance’s memory (a name identifies a series) • After adding a return to the main window of the differential weighing mode SELECTING a series: Press quick access key . A window is displayed with database of series.
Weighing A On selecting an option, the balance moves to the first sample on the list, for which weighing is still to be carried out. If there are no such samples in a series, then the balance orders entering a name of a new sample. Enter the name and accept it by pressing button that is followed by displaying windows for entering the values of ambient conditions (humidity and temperature).
Weighing T Selecting this option opens a screen with keyboard for entering name of a sample. Enter the name and accept it by pressing button that is followed by displaying windows for entering the values of ambient conditions (humidity and temperature). Caution: If the option is disabled in the working mode settings – set to then the software skips the need to specify the ambient conditions parameters.
Weighing T+ A Selecting this option opens a screen with keyboard for entering name of a sample. Enter the name and accept it by pressing button that is followed by displaying windows for entering the values of ambient conditions (humidity and temperature). Caution: If the option is disabled in the working mode settings – set to then the software skips the need to specify the ambient conditions parameters.
In case of a single series, and while weighing the following samples, the measurements can be carried out according to the following outline: weighing A, weighing T or weighing T+A. Each of samples saved in a series shall have a description on its data, i.e. a status, which informs a user on current process phase completed for an individual sample. On entering the database of s and previewing a series, the software opens a screen with saved samples in a selected series.
In case a given series includes samples for which weighing of initial mass (Weighing A) is already completed, it is possible to carry out final weighing (weighing B) for these samples. In such case follow below procedure: Select option: After selecting the option, the balance displays the main window, and the workspace area provides new data on an initiated process.
> pictogram On completing measurements of final mass for the samples in a a user can check the results in the database of s. To check the results, enter the database of s and select a . Select a sample to be previewed. The opened window provides the following information: displayed next to weighing data denote measuring cycles. Data on a specific can be printed or exported to a file.
18.17.3. Copy tare The function enables copying a selected tare value of a sample in a series to all samples that do not have a tare assigned and for which the differential weighing process (status other than Weighing B) is not completed. Means of operation: Press button To open window with a list of samples for which the tares are assigned. Press the tare value field to be applied to other samples in a (those not having a tare assigned).
18.17.5. Deleting a value The function, enables deleting last weighment added to the database, in case an error occurred. It applies to all processes in the working mode. The software enables deleting only the last saved mass measuring record. On accidental saving a weighment (sample or tare), and if the process is accidentally accepted (i.e. wrong sample of container), press the following button: The weighments is automatically deleted and the software returns to the previous step of the procedure.
19. ACCESS TO BALANCE PARAMETERS Balance 3Y series enables determining access level for each of user PROFILES. The user profiles are accessed by: Pressing a field with user name Pressing F1 button on balance overlay Pressing SETUP pictogram or SETUP button followed by pressing Users pictogram The access level limits the possibility of: Modifying balance settings Carrying out balance adjustment process Each user can have other access level assigned.
19.1.1. Assigning password to an administrator PROCEDURE: o Press field with name of a user, or press F1 button o Press field named ADMINISTRATOR o To open a window with list of available settings on a user o Press USER field o The screen displays options on: − User name − Password − Language − Assigning access level − Printing settings − Copying settings o Press field on password setting and using an on-screen alphanumeric keyboard enter a password authorizing access to balance settings.
20. TECHNICAL DATA 20.1. Balance MYA .3Y series Adjustment Display Interface Working temp. Power supply Sensitivity drift Max humidity Min humidity internal 5, 7” colourful touch screen display 2 x USB, 2 x RS 232, Ethernet + 10o C - + 40o C feeder 100V-240V , 50-60 Hz / 13,5-16V DC 1x10-6/ºCxRt 65 % 45 % MYA 2.3Y MYA 0,8/3.3Y MYA 5.3Y MYA 11.3Y MYA 21.
APPENDIX A – cooperation with EPSON printer An EPSON printer is a device subordinate to a balance, thus any data is sent from balance level. Correct cooperation between a balance and a printer requires setting the baud rate to 9600bit/sec. as standard the baud rate of a balance is set to 4800 bit/sec. If a printer cooperates with the balance via its USB port, then the baud rate settings have no relevance.
Cable: balance – printer (CITIZEN, EPSON) Caution: It is recommended to use peripheral equipment supplied by the manufacturer. APPENDIX B – cooperation with KAFKA printer An KAFKA printer is a device subordinate to a balance, thus any data is sent from balance level. Correct cooperation between a balance and a printer requires setting the same baud rate for both devices. Standard balance baud rate is set to 4800 bit/sec. The baud rate setting in a KAFKA printer is carried out using micro switches.
Cable connecting a balance and a printer In case of using the USB port use a standard cable for connecting the devices. When using the RS 232 interface, use cable DB 9/F – DB 25/M. Cable cross-section: Cable: balance – printer KAFKA Caution: - It is recommended to use peripheral equipment supplied by the manufacturer. - a printer KAFKA SQS series is a statistical one featuring two working modes: statistics and txt.
o Set width and height of a label and distance between the labels. o The port settings have no relevance as the printing process is carried out on a connection between a balance and a printer. o Check printer settings regarding paper type and print intensity. o If a label is printed from a computer, check printer settings and port settings. Adding constant texts o Press pictogram and enter a text to be printed. Text size is adjustable. Other fields not related to texts are not relevant.
o The look of a label and quantity of fields are determined by a user o A barcode is added by pressing pictogram. After adding fill in numeric value of the code. o Other objects or images are added by pressing or pictograms. Adding graphics o Graphics is added to a label by pressing a pictogram with a graphic.
o Pressing “PLUS” pictogram enables adding constant texts to custom printout templates – if a printout is a label, then select option ADD FROM USB. o In an opened message box enter name for the text to include data on a label , e.g. production Caution: The data on a label is entered a texts. The texts creating a label are editable. o Go to option “PRINTOUTS” o Press “PLUS” pictogram and set a name for a created non-standard printout, e.g. label.
o Now the non-standard printout contains all parameters required for printing a label. Caution: the balance enables storing multiple label templates for printing, however, each template must have an individual name. 3. Phase III – printing a label o Balance menu requires determining the type of printout, i.e. label printout (give name of the non-standard printout template).
APPENDIX D – databases Databases of a balance enable a user to create a list of data on products, clients, product flow process, etc. The databases provide quick and easy access to data from an optional working mode, and keep records of weighments and controls with complete set of data on a carried out process and weighed product. Menu “Databases” enables determining content of individual databases.
Functions available in each of the databases: Database of products: Sort Select Preview Print Export Import Database of operators: List of saved operators Database of weighments: Select Date from Date to Preview Print Export Database of clients: Sort Select Preview Print Export Import Database of controls: Select Name Preview Print Export Database of average tares: Select Name Preview Print Export Database of packages: Sort Select Preview Pri
Export Import Database of labels: Standard printout Non-standard printout Code page Printout cut Database of ID codes: Sort Select Preview Print Export Import Database of formulations: Sort Select Preview Print Export Import Sort – function enables determining the means of displaying data in the database. Depending on function setting (name, code, EAN code) the data is sorted by the enabled criterion.
On an opened on-screen alphanumeric keyboard enter first letter or letters of searched product name and press button. The software automatically displays only those products that names start with the entered letter(s). Press field with name of the searched product. The software returns to the function with automatic selecting a record of the searched product. Preview – the function enables displaying a list of saved records in a database.
EDITING DATA IN A DATABASE Caution: 1. Database editing is enabled after logging in to a balance as an Administrator, 2. Adding entries to a database does not refer to the database of weighments. Depending on the database type, the software allows for using different options, i.e. functions of the database of products are different than the functions of the database of warehouses. Means of entering data to a database is demonstrated on the database of products.
Each product entry in the database of products features the following parameters: Name Price Description CPG mode (Control of Prepackaged Goods) Code Series EAN code Unit Mass Density Min Shelf life time Max Date Tolerance VAT Tare Label DATABASE OF WEIGHMENTS A separate type of database is the database of weighments. It is used for on-going saving and storing all measurement results carried out on a balance.
Previewing data on an individual weighment is carried out by pressing field with its name. the software opens a window with complete details on a selected weighment. Printing a weighment data is carried out as in case of other databases. Determine selection criteria of data to be printed and press buton. The software automatically sends data complying with the set criteria to a printer connected to a balance.
DATABASE OF WAREHOUSES Each warehouse in the database of warehouses features the following parameters: Name Code Description User manual no.
MANUFACTURER OF ELECTRONIC WEIGHING INSTRUMENTS RADWAG WAGI ELEKTRONICZNE POLAND, 26 – 600 Radom, Bracka 28 Phone. +48 48 38 48 800, fax. + 48 48 385 00 10 export@radwag.com www.radwag.