User manual
Table Of Contents
- 1. BASIC INFORMATION
- 2. TRANSPORT AND STORAGE
- 3. UNPACKING, INSTALLATION AND COMMISSIONING
- 4. GETTING STARTED
- 5. BALANCE LEVELING
- 6. MENU STRUCTURE
- 7. WEIGHING
- 8. BASICS ON TERMINAL AND SOFTWARE OEPRATION
- 9. BALANCE SOFTWARE
- 10. BALANCE ADJUSTMENT
- 11. DISPLAY CALIBRATION
- 12. BALANCE USER SELECTION
- 13. PERIPHERALS
- 14. ABOUT
- 15. COMMUNICATION
- 16. UPDATE
- 17. LEVELING
- 18. BALANCE ADAPTATION – USER SETTINGS
- 18.1. Adding a new operator
- 18.2. Deleting a operator
- 18.3. Selecting a operator
- 18.4. Operator settings
- 18.5. Quick access to determining working modes parameters
- 18.6. Parameter of working mode WEIGHING – quick access keys
- 18.7. Parameter of working mode WEIGHING – information
- 18.8. Parts Counting
- 18.9. Checkweighing
- 18.10. Dosing
- 18.11. Percent setup
- 18.12. Animal weighing
- 18.13. Density of solids, liquid and air
- 18.14. Formulations
- 18.15. Statistics
- 18.16. Mass Comparison
- 18.17. Differential weighing
- 19. ACCESS TO BALANCE PARAMETERS
- 20. TECHNICAL DATA
- APPENDIX A – cooperation with EPSON printer
- APPENDIX B – cooperation with KAFKA printer
- APPENDIX C – cooperation with label printer CITIZEN
- APPENDIX D – databases

- 100 -
18.17. Differential weighing
Differential weighing mode enables analyzing changes of mass in a
single sample or multiple samples.
The process is carried out by determining sample’s initial mass, and
then the sample is subject to different processes, which outcome is
separating or adding of some of sample’s ingredients to its initial
status.
Finally, the sample is repeatedly weighed (differential weighing). After
the final weighing, the balance determines the difference between the
two mass values (initial and final).
Balance 3Y series, in differential weighing mode enables:
Determining up to 200 measuring series, where each series may contain
several samples (each series may contain maximum 1000 samples, assigning
the samples is carried out automatically by adding the following samples during
weighing process of a selected )
Each series can be given a name, printed and data on a can be exported to an
external data storage device
Each sample requires determining tare value, start mass and carry out up to
five consecutive weighments till determining the end mass
If balance users need to use differential weighing mode, then initially, they need to
introduce data on a weighed to balance’s database and then recall a saved for
weighing. The process of creating a is possible from the internal menu (settings) of
the differential weighing mode. The creating procedure is described further in this
user manual.
The differential weighing mode provides access to other functions, such as:
o Buttons
o Information
o Automatic printout
o Printouts /standard – non-standard/
o Printout button /stable – always - disabled /
o ID definition
o Automatic tare
o Counter
o Tare
o Continuous transmission
o Totalized weighments
o s (defining a )
o Threshold (mass value as a maximum mass of a sample, eg. a filter)
o Ambient conditions (if set to <Yes> then the software requires specifying
the temperature and humidity values of ambient conditions before starting
the weighments: <Weighing A, Weighing T+A, Tarring T>)
o Number of repetitions (determining the number of repetitions of the end
weighment <B> – Max no. of repetitions: 5, the settings is valid for all
measured s)