User manual
Table Of Contents
- 1. BASIC INFORMATION
- 2. TRANSPORT AND STORAGE
- 3. UNPACKING, INSTALLATION AND COMMISSIONING
- 4. GETTING STARTED
- 5. BALANCE LEVELING
- 6. MENU STRUCTURE
- 7. WEIGHING
- 8. BASICS ON TERMINAL AND SOFTWARE OEPRATION
- 9. BALANCE SOFTWARE
- 10. BALANCE ADJUSTMENT
- 11. DISPLAY CALIBRATION
- 12. BALANCE USER SELECTION
- 13. PERIPHERALS
- 14. ABOUT
- 15. COMMUNICATION
- 16. UPDATE
- 17. LEVELING
- 18. BALANCE ADAPTATION – USER SETTINGS
- 18.1. Adding a new operator
- 18.2. Deleting a operator
- 18.3. Selecting a operator
- 18.4. Operator settings
- 18.5. Quick access to determining working modes parameters
- 18.6. Parameter of working mode WEIGHING – quick access keys
- 18.7. Parameter of working mode WEIGHING – information
- 18.8. Parts Counting
- 18.9. Checkweighing
- 18.10. Dosing
- 18.11. Percent setup
- 18.12. Animal weighing
- 18.13. Density of solids, liquid and air
- 18.14. Formulations
- 18.15. Statistics
- 18.16. Mass Comparison
- 18.17. Differential weighing
- 19. ACCESS TO BALANCE PARAMETERS
- 20. TECHNICAL DATA
- APPENDIX A – cooperation with EPSON printer
- APPENDIX B – cooperation with KAFKA printer
- APPENDIX C – cooperation with label printer CITIZEN
- APPENDIX D – databases

- 112 -
18.17.3. Copy tare
The function enables copying a selected tare value of a sample in a series to all
samples that do not have a tare assigned and for which the differential weighing
process (status other than Weighing B) is not completed.
Means of operation:
Press button
To open window with a list of samples for
which the tares are assigned.
Press the tare value field to be applied to other
samples in a (those not having a tare
assigned).
The software automatically applies the
selected tare value to data of those samples.
18.17.4. Use of “SAMPLE SELECTION” function.
The function enables
selecting a sample that is prepared for weighing, with no need
to go through all the samples in a series, and for which the process is available. The
function is available only after initiating a process <Weighing A> or <Weighing B>.
Means of operation:
While in the differential weighing mode, and
after selecting a , press <Weighing B> button.
The balance starts the process if a series
contains any samples for which the final mass
measuring process <B> is available.
Press button
To open a list of samples for which the above
procedure is available.
Select a sample to be weighed from the
displayed list.
This option is particularly useful if a series
contains a large number of samples, and the
user should weigh a sample that is at the end
of a series.