User manual
Table Of Contents
- 1. BASIC INFORMATION
 - 2. TRANSPORT AND STORAGE
 - 3. UNPACKING, INSTALLATION AND COMMISSIONING
 - 4. GETTING STARTED
 - 5. BALANCE LEVELING
 - 6. MENU STRUCTURE
 - 7. WEIGHING
 - 8. BASICS ON TERMINAL AND SOFTWARE OEPRATION
 - 9. BALANCE SOFTWARE
 - 10. BALANCE ADJUSTMENT
 - 11. DISPLAY CALIBRATION
 - 12. BALANCE USER SELECTION
 - 13. PERIPHERALS
 - 14. ABOUT
 - 15. COMMUNICATION
 - 16. UPDATE
 - 17. LEVELING
 - 18. BALANCE ADAPTATION – USER SETTINGS
- 18.1. Adding a new operator
 - 18.2. Deleting a operator
 - 18.3. Selecting a operator
 - 18.4. Operator settings
 - 18.5. Quick access to determining working modes parameters
 - 18.6. Parameter of working mode WEIGHING – quick access keys
 - 18.7. Parameter of working mode WEIGHING – information
 - 18.8. Parts Counting
 - 18.9. Checkweighing
 - 18.10. Dosing
 - 18.11. Percent setup
 - 18.12. Animal weighing
 - 18.13. Density of solids, liquid and air
 - 18.14. Formulations
 - 18.15. Statistics
 - 18.16. Mass Comparison
 - 18.17. Differential weighing
 
 - 19. ACCESS TO BALANCE PARAMETERS
 - 20. TECHNICAL DATA
 - APPENDIX A – cooperation with EPSON printer
 - APPENDIX B – cooperation with KAFKA printer
 - APPENDIX C – cooperation with label printer CITIZEN
 - APPENDIX D – databases
 

- 116 - 
20. TECHNICAL DATA 
20.1. Balance MYA .3Y series 
  Adjustment     internal 
  Display    5, 7” colourful touch screen display 
  Interface    2 x USB, 2 x RS 232, Ethernet 
  Working temp.    + 10
o
 C - + 40
o
 C 
  Power supply    feeder 100V-240V , 50-60 Hz / 13,5-16V DC 
  Sensitivity drift    1x10
-6
/ºCxRt 
  Max humidity    65 % 
  Min humidity    45 %  
MYA 2.3Y  MYA 0,8/3.3Y  MYA 5.3Y  MYA 11.3Y  MYA 21.3Y 
Max capacity 
2 g  0,8/3 g  5 g  11 g  21 g 
Reading interval 
1μg  1μg/0,01mg  1μg  1μg  1μg 
Tare range 
-2 g  - 3 g  - 5 g  - 11 g  - 21 g 
Repeatability 
1 μg  1 μg  1,6 μg  2 μg  2 μg 
Linearity 
±3 μg  ±3 μg  ±5 μg  ±6 μg  ±7 μg 
Pan size 
Ø16  Ø16  Ø26  Ø26  Ø26 
MYA 0,8/3.3Y.F  MYA 5.3Y.F  MYA 5.3Y.F1 
Max capacity 
3 g  5 g  5 g 
Reading interval 
1/10 μg  1 μg  1 μg 
Tare range 
- 3 g  - 5 g  - 5 g 
Repeatability 
1 μg  2,1 μg  2,1 μg 
Linearity 
±3 μg  ±5 μg  ±5 μg 
Pan size 
∅ 60  ∅ 100  ∅ 160 
* - in case of any problems with static electricity, it is recommended to increase ambient 
humidity to 50 %. In particular cases of static electricity phenomenon, and if maintaining 
humidity level of approximately 50 % is impossible, use an air ionizer for eliminating 
electrostatic effect on balance indications. 