20. TECHNICAL DATA
20.1. Balance MYA .3Y series
Adjustment internal
Display 5, 7” colourful touch screen display
Interface 2 x USB, 2 x RS 232, Ethernet
Working temp. + 10
o
C - + 40
o
C
Power supply feeder 100V-240V , 50-60 Hz / 13,5-16V DC
Sensitivity drift 1x10
-6
/ºCxRt
Max humidity 65 %
Min humidity 45 %
MYA 2.3Y MYA 0,8/3.3Y MYA 5.3Y MYA 11.3Y MYA 21.3Y
Max capacity
2 g 0,8/3 g 5 g 11 g 21 g
Reading interval
1μg 1μg/0,01mg 1μg 1μg 1μg
Tare range
-2 g - 3 g - 5 g - 11 g - 21 g
Repeatability
1 μg 1 μg 1,6 μg 2 μg 2 μg
Linearity
±3 μg ±3 μg ±5 μg ±6 μg ±7 μg
Pan size
Ø16 Ø16 Ø26 Ø26 Ø26
MYA 0,8/3.3Y.F MYA 5.3Y.F MYA 5.3Y.F1
Max capacity
3 g 5 g 5 g
Reading interval
1/10 μg 1 μg 1 μg
Tare range
- 3 g - 5 g - 5 g
Repeatability
1 μg 2,1 μg 2,1 μg
Linearity
±3 μg ±5 μg ±5 μg
Pan size
60 100 160
* - in case of any problems with static electricity, it is recommended to increase ambient
humidity to 50 %. In particular cases of static electricity phenomenon, and if maintaining
humidity level of approximately 50 % is impossible, use an air ionizer for eliminating
electrostatic effect on balance indications.