User manual
Table Of Contents
- 1. BASIC INFORMATION
 - 2. TRANSPORT AND STORAGE
 - 3. UNPACKING, INSTALLATION AND COMMISSIONING
 - 4. GETTING STARTED
 - 5. BALANCE LEVELING
 - 6. MENU STRUCTURE
 - 7. WEIGHING
 - 8. BASICS ON TERMINAL AND SOFTWARE OEPRATION
 - 9. BALANCE SOFTWARE
 - 10. BALANCE ADJUSTMENT
 - 11. DISPLAY CALIBRATION
 - 12. BALANCE USER SELECTION
 - 13. PERIPHERALS
 - 14. ABOUT
 - 15. COMMUNICATION
 - 16. UPDATE
 - 17. LEVELING
 - 18. BALANCE ADAPTATION – USER SETTINGS
- 18.1. Adding a new operator
 - 18.2. Deleting a operator
 - 18.3. Selecting a operator
 - 18.4. Operator settings
 - 18.5. Quick access to determining working modes parameters
 - 18.6. Parameter of working mode WEIGHING – quick access keys
 - 18.7. Parameter of working mode WEIGHING – information
 - 18.8. Parts Counting
 - 18.9. Checkweighing
 - 18.10. Dosing
 - 18.11. Percent setup
 - 18.12. Animal weighing
 - 18.13. Density of solids, liquid and air
 - 18.14. Formulations
 - 18.15. Statistics
 - 18.16. Mass Comparison
 - 18.17. Differential weighing
 
 - 19. ACCESS TO BALANCE PARAMETERS
 - 20. TECHNICAL DATA
 - APPENDIX A – cooperation with EPSON printer
 - APPENDIX B – cooperation with KAFKA printer
 - APPENDIX C – cooperation with label printer CITIZEN
 - APPENDIX D – databases
 

- 29 - 
10. BALANCE ADJUSTMENT  
Ensuring the highest measuring accuracy of a balance requires 
periodical entering to balance memory a coefficient correcting 
balance indications in relation to a mass standard. The described 
process is known as balance adjustment.  
Adjustment types 
ο  Automatic internal adjustment 
ο  Semi-automatic internal adjustment 
ο
  Adjustment with an external weight / mass standard 
Caution: 
Remember to carry out the adjustment process with no load placed on balance’s weighing 
pan!
Criteria for carrying out the adjustment process: 
  Before starting a weighing process, 
  Between consecutive measuring series, if long breaks occur,  
  In case of considerable change of balance internal temperature. 
The temperature change rate as a factor triggering the adjustment process is 
dependent on balance type, and for the microbalance it is: 
ο  0,4°C – for a balance MYA .3Y series 
If required, balance user can change some of the settings in accordance with below 
description. 
Entering to adjustment settings is carried out by: 
  Pressing SETUP button, or 
  Pressing settings pictogram located in the bottom bar of the 
dispay 
Which opens a window with main software settings. 
Press <Adjustment> field 
to open a drop down window with settings of the adjustment process. 