10. BALANCE ADJUSTMENT
Ensuring the highest measuring accuracy of a balance requires
periodical entering to balance memory a coefficient correcting
balance indications in relation to a mass standard. The described
process is known as balance adjustment.
Adjustment types
ο Automatic internal adjustment
ο Semi-automatic internal adjustment
ο
Adjustment with an external weight / mass standard
Caution:
Remember to carry out the adjustment process with no load placed on balance’s weighing
pan!
Criteria for carrying out the adjustment process:
Before starting a weighing process,
Between consecutive measuring series, if long breaks occur,
In case of considerable change of balance internal temperature.
The temperature change rate as a factor triggering the adjustment process is
dependent on balance type, and for the microbalance it is:
ο 0,4°C for a balance MYA .3Y series
If required, balance user can change some of the settings in accordance with below
description.
Entering to adjustment settings is carried out by:
Pressing SETUP button, or
Pressing settings pictogram located in the bottom bar of the
dispay
Which opens a window with main software settings.
Press <Adjustment> field
to open a drop down window with settings of the adjustment process.