User manual
Table Of Contents
- 1. BASIC INFORMATION
 - 2. TRANSPORT AND STORAGE
 - 3. UNPACKING, INSTALLATION AND COMMISSIONING
 - 4. GETTING STARTED
 - 5. BALANCE LEVELING
 - 6. MENU STRUCTURE
 - 7. WEIGHING
 - 8. BASICS ON TERMINAL AND SOFTWARE OEPRATION
 - 9. BALANCE SOFTWARE
 - 10. BALANCE ADJUSTMENT
 - 11. DISPLAY CALIBRATION
 - 12. BALANCE USER SELECTION
 - 13. PERIPHERALS
 - 14. ABOUT
 - 15. COMMUNICATION
 - 16. UPDATE
 - 17. LEVELING
 - 18. BALANCE ADAPTATION – USER SETTINGS
- 18.1. Adding a new operator
 - 18.2. Deleting a operator
 - 18.3. Selecting a operator
 - 18.4. Operator settings
 - 18.5. Quick access to determining working modes parameters
 - 18.6. Parameter of working mode WEIGHING – quick access keys
 - 18.7. Parameter of working mode WEIGHING – information
 - 18.8. Parts Counting
 - 18.9. Checkweighing
 - 18.10. Dosing
 - 18.11. Percent setup
 - 18.12. Animal weighing
 - 18.13. Density of solids, liquid and air
 - 18.14. Formulations
 - 18.15. Statistics
 - 18.16. Mass Comparison
 - 18.17. Differential weighing
 
 - 19. ACCESS TO BALANCE PARAMETERS
 - 20. TECHNICAL DATA
 - APPENDIX A – cooperation with EPSON printer
 - APPENDIX B – cooperation with KAFKA printer
 - APPENDIX C – cooperation with label printer CITIZEN
 - APPENDIX D – databases
 

- 39 - 
18.4.3. Readout 
Menu readout contains data related to WEIGHING PROCESS, such as:  
  Filter 
Each measuring signal is subject to electronic processing 
before it is displayed. The filtering is carried out to achieving 
correct measurement parameters, specific to a stable signal, 
i.e. one that is ready for reading by a balance user. 
Balance user is able, in a limited range, to change the signal processing 
parameters by selecting one of available FILTER settings. 
Available filter modes: 
o  very fast 
o  fast 
o  normal 
o  slow 
o  very slow 
When selecting the filtering level, take into 
consideration the actual operating conditions of a 
balance. In case of very good operating conditions 
set the filter to normal or fast, in case of rough 
conditions set the filter to slow or very slow.. 
Caution: 
  in case of precision balances 3Y series the recommended range of filter is 
between very fast 
÷
 normal  
  in case of analytical balances and microbalances 3Y series the recommended 
range of filter is between normal 
÷
 very slow 
  Value release 
This parameter determines displaying the stability marker for a 
measurement result. There are 3 available settings of value 
release parameter 
o  fast 
o  fast + reliable / recommended value / 
o  reliable 
Caution: 
The speed of stability marker occurrence depends on applied filter value and applied value 
release.  